3 Remote Operation
3.4.4 Configuration Commands
(continued)
VCO Full‑Scale Frequency
VCOS(?) {
f
}
Set (query) the full‑scale frequency (in Hz) corresponding to a VCO In‑
put voltage of +10 V.
See Section 2.3.1.1 for further discussion the VCO Input.
»
VCOS 5000.0
Example:
Reset
*RST
Reset the SR542 to its default configuration. Sending the
*RST
command
is equivalent to sending the following:
»
MOTR OFF
»
SRCE INT
»
EDGE RISE
»
CTRL OUTER
»
DISP OUTER
»
IFRQ 100.00
»
PHAS 0.0
»
MULT 1
»
DIVR 1
»
VCOS 100.00
»
RELP OFF
»
KCLK ON
»
ALRM ON
SR542 Precision Optical Chopper