Align Refill Switch Mode: The signal that initiates the refill phase
can be displayed during power-up by pressing and holding the PRIME
and the UP-ARROW buttons when the power is switched on. Release
the buttons when the display displays "rFL". When the slotted disk
allows the light beam to pass from the emitter to the detector on the
slotted optical switch a pulse will be generated which signals the
beginning of refill. When this pulse occurs the three horizontal
segments displayed at the top of the display will turn off and the three
horizontal segments at the bottom of the display will turn on. To exit
this mode, press the RUN/STOP button.
Serial Port Loopback Test Mode: If an external device will not
communicate to the pump via the serial port, the serial port loopback
test can be used to verify that the serial port is functioning properly.
During power-up press and hold the UP-ARROW and the DOWN-
ARROW buttons when the power is switched on and then release the
buttons. The display must display "C00" for the first half of the test to
pass. Plug in the serial port loop back plug (A modular plug with pins
2 & 5 jumpered together and pins 3 & 4 jumpered together.). The
display must read "C11" for the second half of the test to pass. To exit
this mode, press the RUN/STOP button.
3.5
Rear Panel Remote Input
An RS-232 modular jacks are provided on the back panel. A
computer, with appropriate software, can be used as a remote
controlling device for pump operation via this connection.
See Appendix A for details on connection and operation.
3.6 Symbols
The following symbols may appear on back panel of the unit:
Caution:
To avoid chemical or electrical hazards, always observe safe
laboratory practices while operating this equipment.
Caution:
To avoid electrical shock and possible injury, remove the power
cord from the back panel of this equipment before performing any
type of service procedures.
Note:
The user shall be made aware that, if equipment is used in a
manner not specified by the manufacturer, the protection provided by
the equipment may be impaired.
3-9
Summary of Contents for BINARY SOLVENT DELIVERY MODULE
Page 2: ......