7-13
10. If storing the pump for more than 12 hours fill each pump head with
100% isopropanol then seal the inlets and outlet with the supplied
plugs.
5.7 Lubrication
The Binary pump has modest lubrication requirements. The
bearings in the pump housing and piston carrier are permanently
lubricated and require no maintenance. A small dab of light grease
such as Lubriplate 630-AA on the cam is the only recommended
lubrication. Be sure not to get lubricant on the body of the piston
carrier, as this can retard its movement and interfere with proper
pumping.
NOTE: Keeping the interior of the pump free of dirt and dust
will extend the pump’s useful life.
5.8
Other Pump Maintenance
The internal components of the pump can be accessed by
removing the pump drawer. To do so, turn off power at front switch.
Unplug power cord.
Ensure there is no power to the system.
Disconnect the inlet line and disconnect pump outlet at the “T”
Connector Block above. Unscrew the two thumbscrews on the upper
portion of the drawer until fully released (counterclockwise). Pull the
drawer out as shown on next page.
Note: Drawer can be removed completely.
Summary of Contents for BINARY SOLVENT DELIVERY MODULE