Recommended separation distances between portable and mobile RF communications
equipment and the Spinal Modulation Neurostimulator System
The Spinal Modulation Neurostimulator System is intended for use in an electromagnetic environment
in which radiated RF disturbances are controlled. The customer or the user of the Spinal Modulation
Neurostimulator System can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the System.
Rated maximum output
power of transmitter
(W)
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.01
0.12m
0.12m
0.23m
0.1
0.37m
0.37m
0.74m
1
1.17m
1.17m
2.33m
10
3.70m
3.70m
7.37m
100
11.70m
11.70m
23.30m
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Conforms to AAMI STD ES60601-1, IEC STDS 60601-1-6, 60601-1-11 &
62366
Certified to CSA STD C22.2 No. 60601-1