14
Figure 11. Connection between the Merlin PCS and the PSA Wand Model 3150 using the PSA
Wand adapter Model 3615
1.
Merlin PCS
2.
PSA Wand Model 3150
3.
PSA Wand adapter Model 3615
PSA Wand Model 3150 Setup
1.
Connect the PSA Wand adapter Model 3615 to the PSA Wand Model 3150.
5
2.
Insert the adapter into the PSA port on the right side or into the inductive telemetry port
inside the storage compartment.
3.
When the Start-Up screen appears, select the Tools button, and then select the PSA button.
The Merlin™ PCS detects the presence of the PSA Wand Model 3150 and starts the
program.
NOTE: The Merlin PCS operates only one pacing system analyzer at a time. If both the
Merlin™ PSA Model EX3100 and the PSA Wand Model 3150 are connected to the
Merlin PCS, the Merlin will only operate the Merlin PSA, and the PSA Wand will not
function.
CAUTION: Do not plug the PSA Wand directly into the PSA port on the right side
of the Merlin PCS without first connecting it to the PSA Wand adapter
Model 3615.
Do not plug the PSA Wand into the PSA port on the Merlin PSA.
RF Communication Setup
Some St. Jude Medical™ implantable devices can communicate with the Merlin™ PCS via radio
frequency (RF), which allows a greater distance between the device and the Merlin PCS. To
establish RF communication, follow the instructions below.
1.
Turn off the Merlin PCS.
5
For information, consult the PSA Model 3150 Reference Manual or press the ? button from the PSA program.