11
RF emissions
CISPR 11
Class B
The PressureWire system is
suitable for use in all establishments
including domestic establishments
and those directly connected to the
public low-voltage power supply
network that supplies building for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
N/A
No mains connection
Voltage fluctuations
/flicker emissions
IEC 61000-3-3
N/A
No mains connection
Guidance and manufacturer’s declaration – electromagnetic immunity
The PressureWire system is intended for use in the electromagnetic environment specified below.
The customer or the user of the PressureWire system should assure that it is used in such an
environment.
Immunity test
IEC 60601
Test level
Compliance level
Electromagnetic environment –
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast transient/
burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
N/A
2)
3)
N/A
Surge
IEC 61000-4-5
±1 kV line(s) to line
(s)
±2 kV line(s) to earth
N/A
2)
N/A
Voltage dips, short
interruptions and voltage
variations to power
supply input lines
IEC 61000-4-11
<5% U
T
(>95% dip in
U
T
) for 0,5 cycle
40% U
T
(60% dip in
U
T
) for 5 cycles
70% U
T
(30% dip in
U
T
) for 25 cycles
<5% U
T
(>95% dip in
U
T
) for 5 sec
N/A
2)
N/A
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
N/A
N/A
N/A
Portable and mobile RF communication
equipment should be used no closer to
any part of the PressureWire system,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
N/A
2)
Recommended separation distance
N/A
2)
No mains power input
3)
No signal or interconnecting cables
Summary of Contents for PressureWire Aeris
Page 2: ...51395139A 4 ...
Page 4: ......
Page 28: ...28 Mettez l émetteur Aeris sous tension ...
Page 143: ...143 Ligar o transmissor Aeris ...
Page 178: ...178 Увімкніть передавач Aeris ...
Page 179: ......
Page 180: ......
Page 181: ......
Page 182: ......
Page 183: ......
Page 184: ......
Page 185: ......
Page 186: ......
Page 187: ......
Page 188: ......
Page 189: ......
Page 190: ......
Page 191: ......