13
Lead Safety
Safety of the Tendril MRI
™
lead was evaluated in terms of freedom from RA and RV lead-related
complications for the acute (implant to two-month visit) and chronic (two-month visit through the
12 month visit) time frames.
MRI Safety
The safety of the Accent MRI system was evaluated in terms of freedom from MRI scan-related
complications in the month following the MRI scan.
Lead Efficacy
Efficacy of the Tendril MRI
™
lead in was evaluated in terms of the change in bipolar atrial and
ventricular capture and sensing thresholds before and after the MRI scan.
Secondary Objectives
The secondary objectives are listed below.
Safety
Safety of the Accent MRI
™
system was evaluated in terms of freedom from system-related
complications through the 12 month visit.
Efficacy
Efficacy of the Tendril MRI
™
lead was evaluated in terms of the bipolar atrial and ventricular
capture thresholds at the MRI Visit.
Patient Selection Criteria
Inclusion Criteria
Eligible patients met all of the following:
1.
Had an approved indication per ACC/AHA/HRS guidelines for implantation of a pacemaker.
2.
Received a new pacemaker and lead.
3.
Was willing to undergo an elective MRI scan without sedation.
4.
Was able to provide informed consent for study participation (legal guardian is NOT
acceptable).
5.
Was willing and able to comply with the prescribed follow-up tests and schedule of
evaluations.
6.
Was not contraindicated for an MRI scan (per the pre-MRI safety screening form)
Exclusion Criteria
Patients were excluded if they met any of the following:
1.
Had an existing pacemaker or ICD. A new pacemaker and lead is required for enrollment.
2.
Had an existing active implanted medical device, e.g., neurostimulator, infusion pump, etc.
3.
Had a non-MRI compatible device or material implanted (e.g., intracranial aneurysm clip,
non-MRI compatible devices or material, metals or alloys, etc.).
4.
Had a lead extender or adaptor.
3
Applies only to those patients who will participate in the MRI Phase of the study
4
Applies only to those patients who will participate in the MRI Phase of the study
5
Applies only to those patients who will participate in the MRI Phase of the study