16
thereafter, during which the following tests/assessments were performed: electrical measurements
on the RA and/or RV leads. Patients were also assessed for adverse events at all study visits.
For the MRI Phase of the study, patients completed an MRI Visit, and returned approximately
thirty days later for a one-month post-MRI visit. At the MRI visit, for patients consenting to
undergo an MRI scan, the following tests/assessments were performed: safety screening for the
MRI scan, the study MRI scan, assessment for adverse events, including MRI scan-related
adverse events, and electrical measurements on the RA and/or RV leads. At the one-month post-
MRI visit, the following tests/assessments were performed: assessment for adverse events,
including MRI scan-related adverse events, and electrical measurements on the RA and/or RV
leads
The table below summarizes all the reported data on the 920 patients who completed the implant
visit.
Table 4. Summary of demographic variables for all enrolled patients
Demographic Variable
All Enrolled Patients (N = 920)
Age
Mean ± SD
73.0±10.8
Range (min, max)
(27,101)
Gender, n (%)
Female
421 (45.8%)
Male
499 (54.2%)
Cardiovascular History, n (%)
Coronary Artery Disease
338 (36.7%)
Myocardial Infarction
119 (12.9%)
Unstable Angina
73 (7.9%)
Prior Cardiac Interventions, n (%)
CABG
130 (14.1%)
PTCA/Stents/Atherectomy
152 (16.5%)
Ablation
100 (10.9%)
Non-Ventricular Arrhythmia History, n (%)
None
370 (40.2%)
AF
481 (52.3%)
Paroxysmal
316 (65.7%)
Permanent
63 (13.1%)
Persistent
100 (20.8%)
AFL
123 (13.4%)
AT
41 (4.5%)
SVT
55 (6.0%)