Vertical check

You must create a reference to perform this test. Fasten a plumbline near to 

the wall. The laser unit should now be set up in front of this reference mark

(distance Y) and the vertical laser line compared with it.

The discrepancy between the centre of the laser line and the reference mark 

should be no greater than 1 mm ( 3/64” ) over a length of 2 m ( 8 ft ). 

Checking plumblining

1. Switch the unit on

2. Place the laser so that its down beam is over a reference mark on the floor.

3. Locate the position of the up beam on the ceiling and make a mark.

4. Rotate the laser 180° and realign the down beam over the reference mark  

    on the floor.

5. Locate the position of the up beam on the ceiling and make a mark.

6.Measure the difference D between the two marks on the ceiling, which will be

   twice the actual error. The difference may not be greater than: 3 mm in 5 m 

    (1/8“ in 16‘5“).

Replacing the batteries

Slide the battery cover (4) in the direction of the arrow and insert new 

batteries in accordance with the symbols in the battery compartment. 

Suitable batteries can also be used.

Technical data

Laser type:     

 

Red diode laser, pulsed line-laser,

   wavelength  635 nm
Output:  

 

< 1 mW, Laser Class 2   to IEC 60825-1:2007

                                                                    This product compl. with the appl. requ. 

                                                               of 21CFR, parts 1040.10 and 1040.11.
Self-levelling range*:   

ca. ± 4,5°

Levelling accuracy :
        laser line horizontal*: 

L1 =  ± 0,3 mm/m     middle of the laser line

                                                                 L1 =  ± 3/16”  over  50ft  
        inclination of the laser line :  L2 =  ± 0,2 mm/m    laser line

                                                              L2 =  ± 1/8”    over  50ft

        Up Beam Accuracy*:   

L3 =   ± 0,3 mm/m  ± 3/16“  over  50 ft

        Down Beam Accuracy:   

L4 =   ± 0,4 mm/m  ± 1/4“    over  50 ft

Batteries::   

 

3 x 1,5 V cells  Alkaline, Size AA, LR6 LR6

Operating life:   Approx. 20 hours (Alkaline)
Operating temperature range : 

-10 °C  to  +50 °C       /     14°F to +122°F

Storage temperature range: 

-20 °C  to  +60 °C      /     -4°F  to +140°F

Subject to technical modifications.

*  When operated within specified temperature range