background image

Diodes and Continuity Measurement (see figure 5)

Warning

To avoid damage to the meter or to the device under test, 
disconnect circuit power and discharge all the high-voltage 
capacitors before diodes.

Use the diode test to check diodes, transistors, and other 
semiconductor devices. The diodes test sends a current 
through the semiconductor junction, and then measures 
the voltage drop across the junction. A good silicon junction 
drops between 0.5V and 0.8V.

To test a diode out of a circuit, connect the meter as follows:

1.  Insert the red test lead into the VΩmA terminal and the 

black test lead into the COM terminal.

2.  Set the knob to Diodes.
3.  For forward voltage drop readings on any semiconductor 

component, place the red test lead on the component’s anode 
and place the black test lead on the component’s cathode.

4.  The measured value shows on the display.

Note

1.  In a circuit, a good diode still produces a forward voltage 

reading of 0.5V to 0.8V; however, the reverse voltage drop 

reading can vary depending on the resistance of other 
pathways between the probe tips.

2.  Connect the test leads to the proper terminals as outlined 

above to avoid error display.

3.  The LCD will display “1” indicating open-circuit for wrong 

connection. The unit of diode is Volt (V), displaying the 
positive-connection voltage-drop value.

4.  When diode testing has been completed, disconnect the 

connection between the testing leads and the circuit test.

General Service

1.  Periodically wipe the case with a damp cloth and mild detergent. Do not use abrasive materials or solvents.
2.  Clean the terminals with a cotton swab with detergent, as dirt or moisture in the terminals can affect readings.
3.  Turn the meter to OFF position when it is not in use and take out the battery when not using for a long time.

4.  Do not store the meter in a place of humidity, high temperature, explosives, flammables, or strong magnetic field.

Replacing the Battery

Warning

To avoid false reading, which could lead to possible electric shock or personal injury, replace the battery as soon as the 
battery indicator        appears. 

To replace the battery:

1.  Disconnect the connection between the testing leads and the circuit under test, and remove the testing leads  

away from the input terminals of the Meter.

2.  Turn the meter to OFF position.
3.  Remove the battery cover.
4.  Remove the battery from the battery compartment.
5.  Replace the battery with a new 9V battery (NEDA 1604 or 6F22 or 006P).
6.  Reinstall the battery cover.

Maintenance

1.  Disconnect the appliance before cleaning.
2.  Use a soft cloth and mild detergent to clean the surface of the meter. To prevent accidental damage or corrosion to the 

surface, do not use an abrasive solvent. 

3.  If you won’t be using this appliance for a period of time, store it in a dry and well ventilated place. Keep it out of reach of children.

Disposal

Environmental Protection

Waste electrical production should not be disposed of with household waste. Please recycle where facilities exist. 
Check with your local authority or retailer for recycling advice.

10

Figure 5

Summary of Contents for DSTS

Page 1: ...Model DSTS User Manual Variable Temperature Digital Solder and Testing Station...

Page 2: ...ll reach as high as 400 C 752 F do not touch metallic parts near the soldering tip when the power switch is on 7 A fire may result if the appliance is not used with care therefore do not leave the app...

Page 3: ...d probe for most testing applications except high amperage 7 Negative socket for black probe Power Supply 8 Coarse adjustment voltage control 9 Negative wire jack 10 Ground jack if needed 11 Positive...

Page 4: ...y elements will be switched on briefly The electronic system will automatically turn on to the set temperature and display the current value Display and temperature setting 1 Display for actual temper...

Page 5: ...er or electrical specifications before using the meter 3 Do not apply more than the rated voltage as marked on the multimeter between the terminals or between any terminal and grounding 4 The rotary s...

Page 6: ...play 2 Display backlight button 3 HOLD button 4 Function selector knob 5 Positive 10A socket for red probe for high amperage testing only 6 Positive socket for red probe for most testing applications...

Page 7: ...000m 10 Battery Type One 9V Battery NEDA 1604 6F22 or 006P 11 Battery Deficiency Display 12 Negative reading Display 13 Overload Display 1 14 Accuracy a reading b digits see charts below guarantee for...

Page 8: ...voltage measurement has been completed disconnect the connection between the testing leads and the circuit under test AC Voltage Measurement see figure 2 Warning To avoid injury to you or damage to th...

Page 9: ...ined 2 When current measurement has been completed disconnect the connection between the testing leads and the circuit under test Measuring Resistance see Figure 4 Warning To avoid damage to the multi...

Page 10: ...t test General Service 1 Periodically wipe the case with a damp cloth and mild detergent Do not use abrasive materials or solvents 2 Clean the terminals with a cotton swab with detergent as dirt or mo...

Page 11: ...and or installation 2 This warranty does not cover cosmetic damage due to misuse or neglect This includes paint damage scratches cracks or other superficial marks related to improper use 3 Failures ar...

Page 12: ...Stahl Tools Last Revised 9 15 2017 stahltools com...

Reviews: