11SOLO2 User’s Manual Revision 0.9
13
Display Scale
Settings
Ctrl ?
Wavelength
Save Settings
Corrections
None
Data Sampling
Multiplier#1 (0.1200000)
Period Multiplier
Offset #1
Trig Level (2%)
Multiplier #2
Refer Values
Offset #2
Fig. 2-1 The menu bar
Some menu options require additional information to be activated properly. In this case, a dialog box
(see Figure 2-2) appears. There you enter the setting values of your choice. Use the
Arrow
keys to
highlight a number and press the
↵
Enter
key to select it. The selected number will appear in the upper
box. Repeat the operation until the desired number has been fully entered. Then go to “OK” and press
↵
Enter
to validate your number. You can correct a mistake by selecting the
Back
button (
←
) located in
the bottom right of the dialog menu. Use the
Decimal
button (
.
) to enter decimals. The
Percent
button
(%) gives you the flexibility to enter a value as a percent rather than a fraction. For example, you can
enter the same multiplier as 0.925 or 92.5%.
Sensitivity (V/J)
200.00
7 8 9
OK
X
4 5 6 %
1 2 3 +/-
0
.
←
Fig. 2-2 The dialog box
2.2 Quick power and energy measurement procedure
This section applies to all 11SOLO2 versions. It will show you the fastest way of making a laser power
and energy measurement with the 11SOLO2 and a Standa power or energy detector.
The monitor automatically recognizes all the Standa power heads and energy heads of version 4 or
higher. All custom technical data required for optimum operation of the detector will be automatically
downloaded from the EEPROM in the DB-15 connector. This data includes sensitivity, model, serial
number, version, wavelength correction factors, and time response. The 11SOLO2 doesn’t support
energy detectors before version 4.
Standa Ltd. All rights reserved