11SOLO2 User’s Manual Revision 0.9
51
Sample Rate
Integer value
Times per unit
Sample Rate Unit
0 = Seconds
1 = Minutes
2 = Hours
3 = Days
4 = percentage of pulses (energy only)
Sample Period
Integer value
Sample Period Unit
0 = Seconds
1 = Minutes
2 = Hours
3 = Days
4 = Weeks
5 = Pulses (energy only)
At the end of a period,
statistics are reset.
Total Duration
Integer value
Total Duration Unit
0
= Sample Period(s)
1
= Seconds
2
= Minutes
3
= Hours
4
= Days
5
= Weeks
6
= Continuous
7
= Predefined number of points
May be a fixed amount of
time or points, one or
many periods, or non-
stop.
Timestamp
0
= Off
1
= On
Adds a time stamp for all
logged data in the
sample.
The following example sets a single data acquisition run for 90 minutes with timestamps at 4 hertz.
Example
Response from 11SOLO2
Text Command:
*DSU 4 0 90 1 1 0 1 <enter>
ACK <CR> <LF>
Binary Command:
Handle = 1152
ACK
Parameter1 = 4
Parameter2 = 0
Parameter3 = 90
Parameter4 = 1
Parameter5 = 1
Parameter6 = 0
Parameter7 = 1
Set Trigger Level
This command sets the internal trigger level when using the device in energy reading mode.
Standa Ltd. All rights reserved