11SOLO2 User’s Manual Revision 0.9
54
Version Extension
2
4 byte word
Length of Device name string
4
4 byte word
Device name string
11SOLO2 Text string in
ASCII code
For this one example we show the binary response as the computer would see it.
Example
Response from 11SOLO2
Text Command:
*VER <enter>
SOLO2 Version 1.01 <CR> <LF>
Binary Command:
Handle = 1001
0 10 0 16 0 0 0 1 0 0 0 2 0 0 0 4 83 79 76
79
Translation
:
Header:
0 10 0 16
0 10 = ACK
0 16 = 16 bytes following header
Reply parameters:
0 0 0 1 0 0 0 2 0 0 0 4
83 79 76 79
0 0 0 1 = version 1
0 0 0 2 = extension .2
0 0 0 4 = 4 characters coming
83 79 76 79 = ASCII code for “SOLO2”
Query Head Name
This command is used to query the model name of the current head.
Text Command
Binary Command
Parameters
Return
HEA
1100
Name of the current heads
The following example is an 11QE-25-SP-MB joulemeter. There are four parameters in binary
mode. In this case they would be:
Binary
parameters Example
Header:
Acknowledge
10
2 byte word
Length of transmission (excluding header)
11
2 byte word
Reply parameters:
Length of Head name string
10
4 byte word
Detector model name
QE-25-
SP-MB
10 bytes
Standa Ltd. All rights reserved