11SOLO2 User’s Manual Revision 0.9
55
Example
Response from 11SOLO2
Text Command:
*HEA <enter>
QE-25-SP-MB
<CR> <LF>
Binary Command:
Handle = 1100
10
QE-25-SP-MB
Query Status
This command is used to view data that is relevant to currently used Detector Head on channel A
or
B.
Text Command
Binary Command
Parameters
Return
STA
1005
Current static and dynamic configuration
values
Configuration values:
Field name
Text Mode Data
Binary Mode Data
1 Head
Type
Indicates whether the current
detector is a wattmeter, a
joulemeter or a photodiode.
1 = wattmeter
2 = joulemeter
3 = photodiode
2 Head
Version
Number identifying the version of the
detector head
same
3 Head Serial Number
Alphanumeric string indicating the
serial number for the head.
Size indicator. The Serial number
string is located at the end of the
binary data.
4 Calibration
Sensitivity
Sensitivity at default wavelength, in
V/W
This is specified in V/W for the Default
Wavelength. Divide this value by
100000. ¹
5 Default
Wavelength
Default Wavelength in nm
same
6 Active
Sensitivity
Currently used sensitivity, it may
have been modified by variables
such as the Active Wavelength
This is specified in V/W for the Active
Wavelength. Divide this value by
100000. ¹
7 Active
Wavelength
Currently used Wavelength in nm
same
8 Scale Min Power
(wattmeters) OR
Max Power
(joulemeters)
Wattmeters: minimum scale index
(see Set Scale command).
Joulemeters: Maximum power
supported by the detector (in Watts).
same
Standa Ltd. All rights reserved