11SOLO2 User’s Manual Revision 0.9
60
3.4.4 ERROR
MESSAGES
# Error
Comment
1
Command not found
Command is invalid
2 Invalid
Parameter
The parameter value is out of valid range, or not of
expected type (text, numeric, flag)
3
Not Enough Parameters
The expected number of parameters should always be
sent.
4
Head is not available
Verify that the detector head is well connected to the
device.
5
Channel is not available
Verify that the detector head is well connected to the
device and that it is recognized by the device.
6
Scale setting not available for specified head
Refer to the Scale table in the Set Scale command
7
Specified channel cannot be offset
Verify that the current detector is in a mode that can
be offset.
8
No Data available
The current configuration cannot provide the
requested data.
9
External trigger is not available
Verify that the device supports external triggering, and
that the current detector is correctly configured.
10 Analog output is not available with External
Trigger
Cannot use the same connection for output and
trigger at the same time.
11 Anticipation is not available
Make sure that the detector is in power-reading mode
12 Statistics are not available
Statistics must first be enabled.
13 PWC is not available
Make sure that the detector head version supports
Personal Wavelength Correction
14 Invalid Command: Too long.
The command must not be over 255 characters long.
15 Too many Parameters
The correct number of parameters must be sent to the
device.
16 Invalid Baud Rate
Verify that the device supports the selected baud rate.
17 Internal trigger is not available
Verify that the detector is configured in a mode that
supports triggering.
18 Calorimeter mode is not available with current
head
Calorimeter Mode works only with some power
meters.
19
Statistics are already enabled
Disable before re-enabling
20 No storage space left
Remove files from the storage media before logging
new ones
21 Head is not a Joulemeter
Make sure the head is a joulemeter before issuing
joulemeter specific commands
22 Already logging data
Disable before re-enabling
23 File does not exist
In order to download a file, first run an acquisition, or
load it from the storage media.
Standa Ltd. All rights reserved