4.6.1. Zero position adjustment
Controller supports setting of zero position. This function should be used for anchor marked stages, so that anchor position matches
logical zero. Also, this function is convenient to use in case there is a single chosen physical position.
To set zero position a
special command
is used. This will zero step/microstep position and encoder count values. Setting of zero
position happens simultaneously for all position counters and will not lead to their desynchronization. Current movement command is
not affected. If controller was moving to some physical position when logical position was reset to zero by this command then the
movement will still end in that physical position. For example, if the controller moved towards logical position 1000 and set zero
position command was sent when it was passing 200, then logical position counter will be decremented by 200 and movement will end
in logical position 800.
Note. Setting of zero position when using shift on offset (see
) will not change target
physical position. Next shift will happen towards the same physical position which would have happened without zero
position command.
Page 105 / 345
Page 105 / 345
Summary of Contents for 8SMC5-USB
Page 30: ...Linux Page 30 345 Page 30 345 ...
Page 83: ...Connection of magnetic brake to one axis or two axis systems Page 83 345 Page 83 345 ...
Page 87: ...Scheme of buttons connection to the HDB 26 connector Page 87 345 Page 87 345 ...
Page 99: ...Page 99 345 Page 99 345 ...
Page 167: ...Vertical load capacity the maximum vertical load on the stage Page 167 345 Page 167 345 ...
Page 186: ...Click Next Click Continue anyway Page 186 345 Page 186 345 ...
Page 188: ...Page 188 345 Page 188 345 ...
Page 190: ...Page 190 345 Page 190 345 ...
Page 193: ...Page 193 345 Page 193 345 ...
Page 201: ...Select the XILab application in the Programs block Start it Page 201 345 Page 201 345 ...
Page 274: ...7 Files 1 Configuration files 2 Software Page 274 345 Page 274 345 ...
Page 331: ...Motorized Iris Diaphragms 8MID98 Motorized Iris Diaphragm Page 331 345 Page 331 345 ...
Page 338: ...8 Related products 1 Ethernet adapter Page 338 345 Page 338 345 ...
Page 339: ...8 1 Control via Ethernet 1 Overview 2 Administration Page 339 345 Page 339 345 ...