4. Operation and application 

a) Connection 

Make sure that you are using a 230V/50-60Hz line voltage. Then plug in the power plug, turn the control 
knobs right (0V / 0A position)  and switch the unit on. Chose the voltage your need.  You can set-up the 
current by switching the Imax switch. When outputcurrent reach the preseted value, mode will change 
into constant current. 

 
Caution! 
 

Parts of the unit ( heatsinks ) will be hot when maximum current is used. The unit has a build in 
overtemperatur protection, which will switch the unit temporally off when 90°C are reached. After a period 
of cooling the unit automatically switches on.  

       Caution! 

Do not cover the ventilation holes. You must ensure that there are at least two inch's clearance around 
the unit.  Take care that the cables of your load are tight connected with the unit. Even low voltage can 
cause electric sparkles that can harm the terminal and your load. 

b) setting up the voltage  (with no load) 

Turn the coarse knob for the current a little clock wise to ensure that the current regulator is not set to 
zero. The LED for constant current is off and the LED for constant voltage is on then. Now you can set up 
your voltage.  

c) setting up the current 

Push the switch Imax. The current display now shows the value for the maximum current you can use. 
Set up the current and release the switch. The unit will now change in to constant current mode if the 
current through your load is reaching se value you set up. 

caution! 

 
Pay attention on the safety instructions in chapter 2 of the manual.. 

 5 specifications 
 
common to the serie 

 

 

ac line voltage 

:  230 VAC +6/-10% 

ac line frequency range 

:  50 Hz 

protection class (IEC 348) 

:   

Common Mode Voltage 

3750 Veff

 from line to terminals

 

Common Mode Voltage 

500V

eff

 

from terminals

 

to ground

 

Warmup time 

:  15 min 

Environmental 

Operating Temperature 

:  + 0 °C bis +35 °C 

Relative Humidity 

(non-condensing)

 

:  max. 85 % by 30 °C 

Air pressure 

:  800 bis 1333 hPa 

 

Summary of Contents for 3231.10

Page 1: ...y to fuses disposable batteries or to any product that has been misused altered neglected or damaged by accident or abnormal conditions of operation or handling We will not be liable for any indirect special incidental or consequential damages Table of contents 1 Introduction 2 Safety instructions connection to line changing of fuses 3 Control Elements Display and Connections 4 Operation and appli...

Page 2: ...tus 2 3 This power supply is for only use on 230V 50Hz line voltage Make sure that the line has this voltage 2 4 Do not leave the power supply in uncontrolled conditions 2 5 Plug off the line cable before you opening of the case Changing fuses and modifying line cables to local power must be done by qualified service personnel who are aware of the hazards involved If you suspect the supply has fai...

Page 3: ...ronic circuits do not mix well with water chemicals dust or grid Avoid exposing your instrument to these or other contaminants The case is not water or air tight Do not place liquid containers on or near the unit where they can be splled into it 2 15 A series connection of several units is allowed only to a overall voltages of 1000V DC 3 Control elements 1 Voltage Display 2 Current Display 3 Switc...

Page 4: ...sparkles that can harm the terminal and your load b setting up the voltage with no load Turn the coarse knob for the current a little clock wise to ensure that the current regulator is not set to zero The LED for constant current is off and the LED for constant voltage is on then Now you can set up your voltage c setting up the current Push the switch Imax The current display now shows the value f...

Reviews: