background image

D2394-01-02 CN4435 DM640P & TC User Guide.doc

1.0 DESCRIPTION

The DM640 series is a range of battery powered temperature indicators.
The DM640P accepts an RTD sensor.
The DM640TC accepts a thermocouple sensor.

The sensor type, temperature units and other flexible configuration settings are
selectable via a simple to use menu system, which is navigated by the use of three push-
button keys located on the rear of the display.  This is described over the page in detail.

The entire assembly is sealed into a cap that fits directly onto the SCH4 series of
connecting heads. Please refer to the SCH4 data sheet for further information.
ATEX approved versions will be available for hazardous, dust and gas areas.

2.0 RECEIVING AND UNPACKING

Please inspect the packaging and instrument thoroughly for any signs of transit damage.
If the instrument has been damaged, please notify your supplier immediately.

3.0 SPECIFICATION @ 20 °C

DM640P

Accuracy

± 0.2 °C ± 0.1% of reading (plus sensor error)

Resolution

0.1 °C (User configurable)

Measuring range

(-100 to +800) °C

Stability 

Zero 0.01 °C / °C
Span 0.005 %/°C

Sensor 

3 wire Pt100 to BS EN 60751(adjustable), or Ni120

Lead effect

10 

 per leg insignificant effect

Units

°C (default) or °F

Sensor type

Pt100 (default) or Ni120

DM640TC

Accuracy

± 0.1 % of FS ± 0.5 °C (plus sensor error)

Resolution

0.1 °C (User configurable)

Measuring range

See table

Stability 

Zero 0.02 °C / °C
Span 0.01 %/°C

Cold Junction Tracking

± 0.05 °C / °C

Units

°C (default) or °F

Sensor type

K (default),J,T,R,S,E,L,N,B

Sensor

Range (°C)

K

-200 to 1370

J

-100 to 1200

T

-210 to 400

R*

-10 to 1760

S*

-10 to 1760

E

-200 to 1000

L

-100 to 600

N

-180 to 1300

B*

0 to 1800

* For type R, S and B sensors, accuracy only applies between 800 and 1760°C

DM640 General

Display

4 digit LCD

Battery Standard

ANSI AA ER14505: IEC ER6

Dimensions

Diameter 14.5 mm, length 50.5 mm

Type

3.6 V Lithium Thionyl Chloride (2.4 Ahr)

Operating Current

80 

µ

A average

Battery Life

> 1 year @ 20 

°

 C

Ambient Temperature

(-10 to 70) °C [Storage (-20 to 85)°C]

EMC  Approval 

Tested to BS EN 61326

Mechanical

Low profile SCH4 Head ABS, IP67 rating when used with base
unit

Connection

Sensor three way screw terminal block to accept 18-22 AWG
wire

4.0 INSTALLATION AND WIRING

4.1 Configuration

Fit the battery as described in section 4.4, then follow the simple
configuration menus printed on the reverse side of this sheet.  This
procedure can be performed with or without the sensor connected.  Fit the correct units
legend to the front panel face.

4.2 Enclosure

The SCH4 case offers various options for sensor entry. Please refer to the SCH4 data
sheet for further details. The sensor may be fixed directly to the SCH4 housing, or
remotely connected using cable and gland. Cable length must be no greater than 10
metres to comply with CE certification. All external cabling/sensor entries must maintain
IP67 rating.

4.3 Wiring

           DM640P Connection

                    DM640TC Connection

        Use correct compensating wire

4.4 Battery

BATTERY WARNING

 

Fire, explosion and severe burn hazard,  DO NOT recharge, crush,
disassemble, heat above 100 °C (212 °F), incinerate or expose contents to
water.

 

Never short circuit the battery.

 

Disposal of the battery may be regulated by  national and local regulations.
Please follow the instructions of the proper regulator.

To install the battery, first gain access to the battery by unscrewing the SCH4 cap
retaining the nut and removing the cap from the base. To remove the old battery, ease
out the positive end of the battery from the holder, using a screw driver blade. To fit the
new battery, insert battery negative into the spring contact end of the battery holder BH1,
then press the battery into place. Replace the cap and secure the nut. Listed below are
suitable
battery types.

Manufacturer

Part Number

Capacity Ahr

Saft

LS-14500, LS-14500C

2.25/2.75

Xeno Energy

XL-060F

2.4

Maxell

ER6

2.0

Tadiran

Tl-2100

2.1

RS

596-602

2.1

Tekcell

SBAA11-TC

2.2

The DM640 devices monitor their batteries.  When the battery approaches the end of its
lifetime (over 1 year under normal conditions), the display will toggle the message "LO
BAT".

5.0 MAINTENANCE

Apart from battery replacement this equipment requires no user
maintenance.  If re-calibration is required, please contact your supplier for further
information.  Under certain circumstances, it may be possible to
re-calibrate the device while it is in an application.

6.0 GENERAL RECOMMENDATIONS

Please observe the battery warning.

Ensure the instrument is installed with adequate protection against the
environment. IP67 rating must be maintained.

Avoid installing the equipment close to sources of extreme temperature and electrical or
electromagnetic interference.

Any cleaning of the instrument should be carried out using a mild detergent and a soft
cloth. No solvents or abrasive cleaner should be used.

7.0 MECHANICAL  DETAIL

The example  below shows an in-head indicator with rotating head
temperature sensor assembly.

8.0 CONFIGURATION MENUS

The two sections overleaf describe the menus accessible on (1) the DM640P, and (2) the
DM640TC.

DM640P & DM640TC

BATTERY POWERED DIGITAL

THERMOMETERS

Designed, manufactured and supported by:

Green Lane, Business Park, Green Lane

Tewkesbury Glos. G20 8DE. UK

Tel: +44 (0)1684 296818    Fax: +44 (0)1684 293746

Email: 

support@status.co.uk

Every effort has been taken to ensure the accuracy of this specification, however we do
not accept responsibility for damage, injury, loss or expense resulting from errors and
omissions, and we reserve the right of amendment without notice

Stock code 52-314-2394-01