18
Electrostatic discharge
IEC 61000-4-2:2008
Contact:
±
8kV
Air:
±
2kV,
±
4kV,
±
8kV,
±
15kV
Contact:
±
8kV
Air:
±
8kV,
±
15kV
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Electrical fast transient/ burst
IEC 61000-4-4:2012
Input AC power ports:
±
2kV
Input AC power ports:
0.5kV, 1kV, 2kV
Mains power quality should be that of a
typical
commercial
or
hospital
environment.
Surge
IEC 61000-4-5:2014
Input AC power ports:
±
0.5kV,
±
1kV,
±
2kV
Input AC power ports:
0.5kV, 1kV
Mains power quality should be that of a
typical
commercial
or
hospital
environment.
Power frequency magnetic
field
IEC 61000-4-8:2009
50Hz or 60Hz
30A/m (rms)
50Hz or 60Hz
30A/m (rms)
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or hospital
environment.
Voltage dips
Voltage interruptions
IEC 61000-4-11:2004 +A1:
2017
PASS
Mains power quality should be that of a
typical
commercial
or
hospital
environment. If the user of
The Dental
Root Canal Orifices Detector
requires
continued operation during power mains
interruptions, it is recommended that
The
Dental Root Canal Orifices Detector
be powered from an uninterruptible power
supply or a battery.
NOTE UT is the a.c. mains voltage prior to application of the test level.