Q-Logic™ Setup
10.0
15
Tremor Suppression
10.4
The
Tremor Suppression
parameter sets a neutral range that suppresses possible tremors (trembling
of hand or drive surface conditions) on the joystick. Setting from 0% to 100%.
▪
Follow
Steps 1 & 2
from
Section 10.2
▪
Select
m: Tremor Suppression
option with the navigation's
Up & Down
buttons, then
press the
Right
button.
▪
Use the
Plus & Minus
buttons to adjust the
Center Deadband
. To turn suppression
off, adjust the setting to 0%.
▪
Press
Left
navigation button
to save.
NOTICE
When short command is used to operate the power chair,
the Tremor Suppression should not be set over 90%, otherwise
any short throw will be ignored.
Assign Direction Function
10.5
The Assign Direction function allows you to set the axis direction. For example, if you want to
change Forward (X axis 0 to 100) to Reverse (X axis 0 to -100) or Left (Y axis 0 to -100) to Right (Y 0
to 100). If you switch Forward to Reverse, the system will automatically change Reverse to Forward.
Same with Left and Right. Parameters are measured in percent (%).
▪
Follow
Steps 1 & 2
from
Section 10.2
▪
Select
Assign Direction
option with the navigation's
Up & Down
buttons, then press
the
Right
button.
▪
Follow the on-screen instructions. When finished, press OK to save.