Error Codes and Fault Logs
14.0
29
Fault:
Reason for Fault:
Required Action:
CPU Error: RAM
CPU consistency check failed
Replace PCB.
CPU Error: FLASH
CPU Error: EEPROM
Run Error: Scheduler
Firmware consistency check failed
- Update software.
- Replace PCB.
Code Error: Framework
Code Error: Application
MSP Command Corrupt
Corrupt Command Received
Connection with PC
(Configuration Software)
went wrong, try again.
MSP Command Unknown
Unknown Command Received
Connection with PC
(Configuration Software)
went wrong.
Update Firmware or
Update Configurator
Software. Try again.
MSP Sub Command
Unknown
Unknown Sub Command Received
MSP Argument Invalid
Invalid Argument Received
MSP Device Not Ready
Device not ready
to receive MSP command
MSP Device Wrong State
Device unable to receive MSP
command in current state
PCB Test Failed
Factory test failed
A fault occurred during
factory testing.
Assembly Test Failed