8.  [Samp] key: Put enough sample on the platform, then input the sample quantity, it 

first displays in unit weight window. Then press this key while the weight display 
is stable, the number shifts to number window, the unit weight window displays 
the unit weight of the sample. If the unit weight calculated is less than 80% of the 
graduation, the “U.W” signal display. If the weight of sample is less than 4 time of 
graduation, then the “sample” signal display. If you need to change the quantity 
of sample , you should first press the key [Clear], then input the new number. 
If you’ve input the number for 3 seconds, then you can input the new number 
directly without pressing the key [Clear].When the unit weight is max. four 
decimal points, the unit weight value can’t exceed 1 kg. Or there will be calculate 
mistake.

9.  Backlight On/Off: Press [Zero] key for 3-4 seconds, the backlight will be switched 

on or off.

10.  Adjust the decimal point number for the unit weight: Enter the unit weight form 

the keyboard, when the unit weight is zero, press number key “0” to adjust the 
decimal point number for the unit weight.

ALARM INDICATION

1.  The weight window displays “-----” with continuously alarm sound when the-

weight over 100%FS+9d, if the ADC overflows, the weight window displays 
“—Adc—”with continuously alarm sound. For these instance, the weight 
should be taken away.

2.  The number window displays “—OF—” when the number or accumulated 

number over 99999. The scale will not response if the accumulated times 
over 100, but the original accumulated data will be retained.

3.  If the voltage of the accumulator is low, the weight window displays “—Lb—” 

while the weight is zero (The weight display will recover to normal when it is 
loaded ). Under this condition, you can use the scale in a short time, but you 
should plug in the AC plug as soon as possible to recharge the accumulator.

4.  The weight window displays “HHHH” or “LLLL” when the zero weight is 

higher or lower than the permitted range.

5.  If the weight window displays “-SYS-”, it means the system parameter is 

wrong., you should send it for maintain immediately.

6.  If the weight window displays “Err-2” after self-testing, it means unstable 

weighing, please see III/1/(2),(3).

WEIGHT CALIBRATION

1.  Press [A.C.] key till the weight window display “-CAL.-”, the unit weight window 

and the number window display 0 or smaller number

2.   Weight calibration:

• 

F.S. calibration: Put enough weight on the platform. press [Accu] when display is 
stable. The calibration finish.

• 

Calibrate by any weight: Press [Tare], then weight window displays “-----“. Unit 
weight window and number window don’t change. Input the weight by keyboard, 
then put the corresponding weight on the platform. Press [Accu] when display is 
stable. The calibration finish.

REGULAR TROUBLESHOOTING

1. Unstable weighing

• 

Low voltage for accumulator: Switch off the scale immediately and recharge it 
more than 12 hours.

• 

Low intensity or strong shake for platform: Strengthen the platform or avoid-
shake.

• 

Influenced by the wind: Set a glass cover to reduce the influence of the wind.

2. No display when switch on the scale

• 

Accumulator is unusable: Change the accumulator if it remains invalid after 
recharging.

• 

Broken with fuse or switch: Change the fuse or switch.

• 

Others: Send it for maintain.

TRANSPORTATION AND STORAGE

• 

Avoid shock and dampness during transportation.

• 

Avoid strong sunshine and rain during storage.

ACCESSORIES

• 

Instruction Manual 

• 

Power supply wire

ATTENTION

1.  Please read this instruction manual carefully before you use the scale.
2.  If it has not be used for long time, please recharge it once 2 months at least, and-

keep the power switch on “off” status to avoid discharge of accumulator.

3.  LCD of the scale must to be operated under rich light, you are advised to select 

product with backlight under low light condition. 

4.   Avoid collision when you put the weighting object on the scale.

EN

EN

Summary of Contents for SBS-ZW 3001H

Page 1: ...BEDIENUNGSANLEITUNG ELEKTRONISCHE ZÄHLWAAGE USER MANUAL ELECTRONIC COUNTING SCALE ISTRUZIONI D USO BILANCIA CONTAPEZZI ELETTRONICA MANUEL D UTILISATION BALANCE ÉLECTRONIQUE DE EN IT FR SBS ZW 3001H ...

Page 2: ...erscheint auf dem Display Bereich 4 FS 3 Tara Taste Tara einstellen oder löschen Bereich 1 3F S Legen Sie den Gegenstand auf die Wiegeplatte anschließend drücken Sie die Taste das Gewichtfenster zeigt Null an Tara leuchtet in der Anzeige auf Wenn die Nullanzeige länger angezeigt wird dann kann die Waage nicht genutzt werden Anderenfalls wird die automatische Nullstellenüberwachung nicht funktionie...

Page 3: ...aage nur kurz genutzt werden und sie sollte schnellstmöglich an die Stromversorgung ang eschlossen werden um den Akku aufzuladen 4 Das Gewichtfenster zeigt HHHH oder LLLL an wenn das Nullgewicht höher oder niedriger als der zulässige Bereich ist 5 Zeigt das Gewichtfenster SYS an dann bedeutet das dass die Systemparam eter falsch sind und umgehend korrigiert werden sollten 6 Zeigt das Gewichtfenste...

Page 4: ...ierungsmodus wechseln drücken Sie die Taste TARE um den entsprech enden Modus zu wählen Es gibt A Kalibrierung mit einem Gewicht B Kalibrierung mit zwei Punkten und C Kali brierung mit drei Punkten sieht wie folgt aus A 1 LOAD für die Kalibrierung mit einem Gewicht drücken Sie die Taste ACCU zur Bestätigung Der Kapazitätswert wird angezeigt zur gleichen Zeit kann das Gewicht über die Tastatur eing...

Page 5: ...ss the key the weight window displays zero the ta resignal displays You can t use the scale in the state of zero signal displaying in longterm otherwise the function of auto zero tracing will disappear and the zero point will excurse Take away the weighing object scale indicate negative press this key again to clear tare 4 0 9 number key For input the data from right to left in turn in Unit Weight...

Page 6: ... to recharge the accumulator 4 The weight window displays HHHH or LLLL when the zero weight is higher or lower than the permitted range 5 If the weight window displays SYS it means the system parameter is wrong you should send it for maintain immediately 6 If the weight window displays Err 2 after self testing it means unstable weighing please see III 1 2 3 WEIGHT CALIBRATION 1 Press A C key till ...

Page 7: ...choose the value bigger the zero track rang will bigger press ACCU zur Bestätigung 4 Calibration mode choosen press TARE key choosen calibration mode in here has A Calibration by any weight B calibration by two point and C calibration by three point Operation as below A 1 LOAD for calibration by any weight press ACCU keyconfirm display the capacity value at this time can input the weight value by ...

Page 8: ...ggiungere lo status di pesata stan dard 2 Tasto zero una volta azzerato il peso il display di pesata indichera il numero zero 4 SF 3 Tasto tara Impostare o cancellare la funzione tara 1 3SF Appoggiate gli articoli sul piatto di pesata e infine premete il tasto tara Sul display di pesata comparira il numero zero e l indicatore della funzione TARA si illumi nera Qualora il numero ZERO permanesse a l...

Page 9: ...ifra digitata o la somma del peso ha supera to il numero 99999 La bilancia non reagisce ai dati somma che superano il valore IT 17 100 mentre i dati sommati originariamente rimangono salvati 3 Se la tensione della batteria e bassa la finestra peso indica Lb e il peso indica Zero Il display mostra peso ritorna alla funzione normale una volta che la batteria e completamente carica In questo caso la ...

Page 10: ...iderata in kg e in g Per continuare la procedura si prega di digitare il numero 090316 e di premere successivamente il tasto ACCU 1 Per scegliere la capienza verranno mostrate queste 3 finestre FULL SCALE XXX xxx sta per capacita Con i tasti cifra potrete imposta re la capacita che piu desiderate ad esempio per una bilancia da 30kg digitare il numero 30 qualora l unita di misura fosse espressa in ...

Page 11: ...incipale premendo il tasto AC Calibrazione semplice Durante ogni pesatura premete il tasto AC per circa 3 secondi Compariranno le seguenti finestre 0 LOAD current capacity La misurazione della capacita attuale e la digitazione del peso di calibrazione avvengono premen do i tasti appositi Si prega di lasciare l articolo sul piatto di pesata fino a quando l indicazione del peso non si sia stabilizza...

Page 12: ...ent enregistrées 1 Touche A C Supprime les données stockées quitte la fonction addition et le symbole addition disparaît Touche Samp Mettez l échantillon sur le plateau de pesée puis entrez le nombre d échantillons et celui ci apparaît alors sur l écran poids Appuyez ensuite sur la touche jusqu à ce que la lecture du poids soit stable L écran poids indique le poids de l échantillon Si le poids cal...

Page 13: ...bien éclairé et de régler le rétroéclairage sur une puissance plus faible 4 posez les objets en douceur sur le plateau de la balance évitez les chocs et les secousses 5 Il est strictement interdit d utiliser des solvants chimiques pour le nettoyage du plateau 6 Nettoyez la balance avec un chiffon sec ou légèrement humide Ne la lavez jamais à l eau 7 N utilisez pas de circuit d alimentation trop fo...

Page 14: ... terminé C 3 LOAD3 pour le calibrage avec trois points appuyez sur la touche ACCU un code AD interne s affiche Appuyez sur la touche ACCU de nouveau le symbole LOAD1 apparaît choisissez 1 3 de la capacité appuyez sur la touche ACCU pour confirmer le symbole LOAD2 s affiche alors choisissez les 2 3 de la capacité appuyez sur la touche ACCU pour valider Lorsque LOAD3 apparaît choisissez la pleine ca...

Page 15: ...GHT vous lisez le poids total des pièces 5 Le mode de comptage est prêt à l emploi Zusatzinformationen Supplementary information Information complémentaire MODO DE CONTAJE 1 Encienda la báscula y espere un momento 2 Coloque en el plato un número fijo de piezas por ejemplo 10 lo más lo mejor 3 Seleccione mediante las teclas 1 2 9 el número de las piezas en la bandeja 4 Pulse el botón sample En la p...

Page 16: ...eży bezwzględnie wyłączyć wagę oraz odłączyć zasilanie wagi z zasilacza sieciowego 3 Demontaż akumulatora kwasowo ołowiowego znajdującego się w wadze a Odkręcić wkręty mocujące pokrywę akumulatora znajdującą się z pod spo dem urządzenia b Zdjąć pokrywę c Odpiąć odlutować przewody końcówek akumulatora d Wyjąć akumulator 4 Akumulator przekazać komórce odpowiedzialnej za utylizację tych materiałów 5 ...

Page 17: ...sträger Kommunen haben hierzu Sammelstellen eingerichtet an denen Altgeräte aus privaten Haushalten ihres Gebietes für Sie kostenfrei entgegengenommen werden Möglicherweise holen die rechtlichen Entsorgungsträger die Altgeräte auch bei den privaten Haushalten ab Bitte informieren Sie sich über Ihren lokalen Abfallkalender oder bei Ihrer Stadt oder Gemeindeverwaltung über die in Ihrem Gebiet zur Ve...

Reviews: