THERMOMETER & THERMOSTAT

 

The bi-metal thermostat is used to adjust the temperature of the  sterilizer. The 
glass thermometer on the front of the unit indicates  the 

sterilizer’s chamber 

temperature. 

PILOT LIGHTS

 

The sterilizer has two (2) pilot lights.  The left light will be on as long  as the time 
switch is in the 

“ON” position.  The right light is on only  while the heating elements 

are operating and heating the sterilization  chamber. 

REPLACEMENT FUSES

 

Models 200 & 300:  Slow Blow type, rated 10A, 250V  Models 200 & 300 
(230V):  Slow Blow type, rated 5A, 250V 
The fuse is located on the right side of the unit, next to the power cord. 

CLEANING AND MAINTENANCE

 

Clean the inside and outside of the sterilizer with a stainless steel  cleaner at least 
once a month.  We recommend Cleaning Kit #201009. When cleaning the inside of 
the chamber, spray cleaner on cloth first  and avoid contact with heating elements.  
Use a wet cloth to wipe away  any cleaner residue and then follow with a dry cloth 
to wipe dry. 
Replace the inner door gasket once a year, or sooner if it becomes  loose.  Order 
gasket # 201108 for Model 200 or #301108 for Model 300.  Replace the thermostat if 
the temperature control becomes erratic or  every 3 years.  Order part #201032 for 
Model 200 or part #301032 for Model 300. 

ENVIRONMENTAL CONDITIONS

 

The Steri-Dent sterilizer is designed for indoor use with the following  conditions: 
•  Temperature Range of 5°C to 40°C (41°F to 104°F) 
•  Maximum Relative Humidity of 80% up to 31°C (88°F). 

Decreasing linearly to 50% at 40°C (104°F). 

•  Pollution Degree 2 applies in accordance with IE 664. 
•  Transient Over-voltage Category II applies. 
•  Supply voltage not to fluctuate more than 10% (+/-12V at 120V, +/-22V at 220V) 
•  Maximum altitude of 2000 m (6562 ft). 

 

Steri-Dent sterilizers are the result of a design that has been in use  for many 
years by dentists, hospitals, and the U.S. armed forces.  This design has been 
refined and improved over the years to make it safe and simple to operate, easy 
to use, and low in cost to operate.  It is user friendly and very reliable, but 
requires the operator to choose  the proper loading, temperature and time 
settings to ensure complete  sterilization. 

 
 

ETL LISTED 
CONFORMS TO UL STD 61010-1  CERTIFIED TO 
CSA STD C22.2 NO. 61010-1  CAN/CSA STD 
C22.2 NO 61010-2-040 

FDA 510(K) K771070 

 

 