OPERATING INSTRUCTIONS
MODELS 200 & 300: Read the entire instruction manual before installation or
operation of the Steri-Dent Dry Heat Sterilizer.
WARNING: NEVER ATTEMPT TO PERFORM ANY ELECTRICAL
TROUBLESHOOTING ADJUSTMENT(S) OR SERVICE(S) UNLESS YOU ARE
A QUALIFIED ELECTRICIAN, ELECTRONICS TECHNICIAN OR FACTORY
TRAINED SERVICE TECHNICIAN -
The protection provided by the equipment may be impaired if the Sterilizer is used
in a manner not specified in this manual.
IMPORTANT SAFEGUARDS
1. Read and understand all instructions.
2. Care must be taken to avoid injury or burns from touching hot parts.
3. Do not operate this unit with a damaged cord.
4. Do not operate this unit if it has been dropped or damaged until it has been
examined by a qualified service technician.
5. Do not let power cord hang over edge of table or counter or touch hot
surfaces.
6. An extension cord should not be used with this unit. The unit should be
plugged directly into a power outlet. Only use a properly grounded
fuse/breaker protected outlet (115V, 50/60 cycles or a 220/240V, 50 cycles).
A separate circuit is recommended.
7. To protect against electrical shock hazard, do not immerse this appliance
in water or other liquids.
8. To avoid electrical shock hazard, do not disassemble this appliance. No user
serviceable parts inside. Call a qualified service technician when service or
repair is required.
9. Do not lift unit by drawer handle. Hold securely by the bottom when lifting or
moving unit.
STERILIZER PROCEDURE
Prepare all instruments for sterilization by using an ultrasonic cleaner, as
recommended by the Centers for Disease Control (CDC).
Thoroughly rinse the instruments and pat dry with a paper towel. The CDC
recommends bagging the instruments prior to sterilizing.
Sterilization bags
specific to dry heat sterilization must be used. Bags are available from Steri-
Dent.
Instruments should be placed in a single
layer and distributed evenly on the
sterilizer’s trays. Instruments should not
touch each other. Load instruments into
sterilizer and pre-heat to temperature
indicated in the TIME/TEMPERATURE
chart. Run the sterilization cycle for the
length of time indicated.
NOTE: Allow time for instruments to reach operating temperature before starting
the time cycle.
TIME/TEMPERATURE CHART
Contents
Preparation
Temperature
Time
Metal instruments,
surgical blades, etc
Thoroughly cleaned and
spread out on trays.
320 Degrees F
160 Degrees C
60 minutes
MAXIMUM LOADING
Model 200 Sterilizer
– 1.3 pounds per tray (includes the weight of the tray)
Model 300 Sterilizer
– 1.4 pounds per tray (includes the weight of the tray)
SPORE TEST STRIPS
The Centers for Disease Control (CDC) recommends the sterilizer be tested
weekly. For Dry Heat Sterilizers, use Bacillus Atrophaeus spore test strips.
POWER SOURCE
Standard units operate on 115 VAC electrical power. Other voltages are available
upon request. Voltage rating is printed on the nameplate located near power cord.
Model 200 Power Rating
– 500 Watts Model 300
Power Rating
– 735 Watts
Steri-Dent Dry Heat Sterilizers* are intended for indoor use in hospitals and dental,
orthodontic and health care facilities. They come equipped with a mechanical
automatic timer, bi-metal thermostat, and thermometer.
MECHANICAL AUTOMATIC TIMER
The timer serves as the ON/OFF switch for the sterilizer. To activate the timer and
turn the sterilizer on, turn the knob clockwise (right) past the number
“10”. The
timer can then be set for the desired time as indicated on the calibrated scale dial.
To set the timer for continuous operation or
“warm up”, turn the knob counter
clockwise (left) to the
“HOLD” position. CAUTION: The sterilizer will remain on
until it has been manually turned off or set for a timed sterilization cycle.
STERILIZER SET UP
(First time only)
Before using the sterilizer for the FIRST time, the thermostat setting must be
checked to ensure the sterilizer will operate at the proper temperature. For 320°F
(160°C), verify t
he number “7” on the thermostat dial is lined up with the black
marking dot located on the right side of the sterilizer. Set the timer to
“HOLD”
position.
After approximately 45 minutes, check the thermometer. It should read between
320°F (160°C) and 330°F (165°C). If not, adjust the thermostat dial and wait
approximately 15 minutes for the temperature to change. NOTE: The thermostat
dial is very sensitive. Decreasing the dial by 1 will drop the temperature
approximately 15°F. Continue to adjust the thermostat as needed until the
thermometer reads between 320 Degrees F (160 Degrees C) and 330 Degrees F
(165 Degrees C). Temperatures may vary during initial heat-up period.
*Steri-Dent sterilizers are non-patient care equipment.
WARRANTY:
Visit
www.cpac.com
for warranty information.