background image

OPERATING INSTRUCTIONS

 

MODELS 200 & 300:  Read the entire instruction manual before installation or 
operation of  the Steri-Dent Dry Heat Sterilizer. 

 

WARNING:  NEVER ATTEMPT TO PERFORM ANY ELECTRICAL 
TROUBLESHOOTING ADJUSTMENT(S) OR SERVICE(S) UNLESS  YOU ARE 
A QUALIFIED ELECTRICIAN, ELECTRONICS TECHNICIAN OR FACTORY 
TRAINED SERVICE TECHNICIAN - 
The protection provided by the equipment may be impaired if the Sterilizer is used 
in a manner not specified in this manual.

 

 

IMPORTANT SAFEGUARDS

 

1.  Read and understand all instructions. 

2.  Care must be taken to avoid injury or burns from touching hot parts. 

3.  Do not operate this unit with a damaged cord. 

4.  Do not operate this unit if it has been dropped or damaged until it has been 

examined by a qualified service technician. 

5.  Do not let power cord hang over edge of table or counter or touch  hot 

surfaces. 

6.  An extension cord should not be used with this unit. The unit  should be 

plugged directly into a power outlet. Only use a properly  grounded 
fuse/breaker protected outlet (115V, 50/60 cycles or a  220/240V, 50 cycles).  
A separate circuit is recommended. 

7.  To protect against electrical shock hazard, do not immerse this  appliance 

in water or other liquids. 

8.  To avoid electrical shock hazard, do not disassemble this appliance.  No user 

serviceable parts inside. Call a qualified service technician when service or 
repair is required. 

9.  Do not lift unit by drawer handle. Hold securely by the bottom when lifting or 

moving unit. 

STERILIZER PROCEDURE

 

Prepare all instruments for sterilization by using an ultrasonic cleaner,  as 
recommended by the Centers for Disease Control (CDC). 
Thoroughly rinse the instruments and pat dry with a paper towel.  The CDC 
recommends bagging the instruments prior to sterilizing.  

Sterilization bags 

specific to dry heat sterilization must be used.  Bags are available from Steri-
Dent.

 

Instruments should be placed in a single 
layer and distributed evenly on  the 
sterilizer’s trays. Instruments should not 
touch each other. Load  instruments into 
sterilizer and pre-heat to temperature 
indicated in the TIME/TEMPERATURE 
chart.  Run the sterilization cycle for the 
length of time indicated. 

NOTE:  Allow time for instruments to reach operating temperature before  starting 
the time cycle. 

TIME/TEMPERATURE CHART 

Contents 

Preparation 

Temperature 

Time 

Metal instruments, 

surgical blades, etc 

Thoroughly cleaned and 

spread out on trays. 

320 Degrees F 

160 Degrees C 

 

60 minutes 

MAXIMUM LOADING

 

Model 200 Sterilizer 

– 1.3 pounds per tray (includes the weight of the  tray) 

Model 300 Sterilizer 

– 1.4 pounds per tray (includes the weight of the  tray) 

SPORE TEST STRIPS

 

The Centers for Disease Control (CDC) recommends the sterilizer  be tested 
weekly. For Dry Heat Sterilizers, use Bacillus Atrophaeus  spore test strips. 

POWER SOURCE

 

Standard units operate on 115 VAC electrical power. Other voltages  are available 
upon request. Voltage rating is printed on the nameplate  located near power cord. 
Model 200 Power Rating 

– 500 Watts  Model 300 

Power Rating 

– 735 Watts 

Steri-Dent Dry Heat Sterilizers* are intended for indoor use in hospitals and dental, 
orthodontic and health care facilities. They come  equipped with a mechanical 
automatic timer, bi-metal thermostat, and  thermometer. 

MECHANICAL AUTOMATIC TIMER

 

The timer serves as the ON/OFF switch for the sterilizer. To activate  the timer and 
turn the sterilizer on, turn the knob clockwise (right) past  the number 

“10”. The 

timer can then be set for the desired time as  indicated on the calibrated scale dial. 

To set the timer for continuous operation or 

“warm up”, turn the knob  counter 

clockwise (left) to the 

“HOLD” position. CAUTION: The sterilizer will remain on 

until it has been manually turned off or set for a  timed sterilization cycle. 

STERILIZER SET UP 

(First time only) 

Before using the sterilizer for the FIRST time, the thermostat setting must be 
checked to ensure the sterilizer will operate at the proper temperature.  For 320°F 
(160°C), verify t

he number “7” on the thermostat dial is lined up with the black 

marking dot located on the right side of the sterilizer.  Set the timer to 

“HOLD” 

position. 

After approximately 45 minutes, check the  thermometer.  It should read between 
320°F (160°C) and 330°F (165°C).  If not, adjust the thermostat dial and wait 
approximately 15 minutes for the temperature to change.  NOTE:  The thermostat 
dial is very sensitive.  Decreasing the dial by 1 will drop the temperature 
approximately 15°F.  Continue to adjust the thermostat as needed until the 
thermometer reads between 320 Degrees F (160 Degrees C) and 330 Degrees F 
(165  Degrees C).  Temperatures may vary during initial heat-up period. 

 

*Steri-Dent sterilizers are non-patient care equipment. 

 

WARRANTY:

 Visit 

www.cpac.com

 for warranty information.

 

Reviews: