9 of 40
Operating Manual:
SU780XLE
SU780XLE-LIT-OM-I |
11.18.19
www.stirlingultracold.com
2.5 – Graphical User Interface
The Graphical User Interface (GUI) is implemented on a touch screen control panel
on the front of the cabinet. The GUI allows the user to:
•
Set freezer control, alarm and communications parameters
•
Provide password protected access to parameter changes
•
Display freezer status
•
Display temperature graph and event log
•
Display diagnostic and service information
•
Display freezer identification information
The following SU780XLE parameters are accessed and modified through
the GUI touch screen (for complete instructions see Section 5).
Accessible Values
Modifiable Values
Current freezer temperature
Temperature setpoint
Temperature history chart
Under temperature limit
Alarm status
Over temperature limit
Event log
Over temperature alarm delay after door opening
Battery information
Audible alarm silence duration
"About" screen (general information)
External contact alarm delay
Service contact information
Door open alarm delay
Door defrost percentage
Password Identification Number (PIN)
Current date
Current time
RTD calibration adjustments
2. Features