STM Instruments Inc Page 8
Figure 5 Fluid Temperature vs. Time
To ensure accurate readings when filling hot or cold fluids
follow the same procedure as stated for a non-carbonated
room temperature fluid with the following additional steps.
Fill the syringe with the hot/cold fluid; securely attach the
syringe to the filling tube of the
eDrometer
then depress the
syringe plunger at a faster rate than if you were filling at
room temperature, empty the syringe in 3-5 seconds. The
reason for filling at a faster rate with hot/cold fluids is how
the internal components distribute the temperature. Wait
for the
eDrometer
to stabilize for an additional 30-45
seconds and then take your reading.
Once filling is complete you may notice a “TL” or “TH”
flashing between the displayed temperatures,
Figure 6
,
depending on the fluid temperature.