4.
Once the trim value is cleared you will go back to the
Clear Trim
screen which will show a Trim value of
0.000 and the current SG,
Figure 17
.
Figure 17 Confirm Trim Cleared
5.
Now to return to the normal display press the Scroll
button twice.
NOTE:
-
If the fluid in the eDrometer differs by more than
0.05 SG of the calculated value the trim will be
rejected.
-
If the fluid temperature at the time of the Trim is
less than 15°C and greater than 25°C the Trim will
be rejected.
-
Only use distilled/de-ionized water for the
Trimming process.