Medium Optical Disk (RW551/RW552) Library
4–20
EK–MOL80–SV. B01
Table 4–3 Hardware Errors and Recovery Procedures (
continued
)
Error Code
Error Description
Recovery Procedures
(hex.)
44
Test drive
•
Look at the micro-move error of
the failure in the error log (under
INFO *, and Hardware Error in
the control panel display).
•
Remove rear panel and run the
Wellness Test, Drive I/O test,
and
Exercise Mechanics test. Note
where the problem occurs.
•
If indicates the drive, change the
drive.
•
If it indicates a picker error,
change the picker.
45
Put mailslot in
•
Change the picker.
•
Change the mailslot.
46
Get mailslot out
•
Change the picker.
•
Change the mailslot
47
Test mailslot
•
Change the picker.
•
Change the mailslot.
48
Rotate mailslot in
•
Cycle power to the jukebox to
initiate a power on test
sequence.
•
Check to see if mailslot rotation
works.
•
If the mailslot rotates in, change
the picker.
•
If the mailslot does not rotate in,
change the mailslot.
49
Rotate mailslot out
•
Cycle power to the jukebox to
initiate a power on test
sequence.
•
Check to see if mailslot rotation
works.
•
If the mailslot rotates in, change
the picker.
•
If the mailslot does not rotate in,
change the mailslot.