25
Table 201
Guidelines and Manufacturer’s Declaration - Electromagnetic Emission
The charger model 094129/11301 DG is designed for operation in an environment as defined below.
The user of the device must make sure that the charger model 094129/11301 DG is operated in such an environment.
Interference emission measurement
Conformity
Electromagnetic environment – guidelines
RF emissions
according to CISPR 11
Conforms to Group 1
The charger model 094129/11301 DG uses RF energy exclusively for its internal
functions. Accordingly, its RF emission is very low and it is unlikely that it will
interfere in other nearby electronic devices.
RF emissions
according to CISPR 11
Conforms to Class B
The charger model 094129/11301 DG is suitable for use in all buildings
(e.g. hospitals and medical practices), including those in living quarters and those
connected directly to a public supply system that also supplies buildings that are
used to live in.
Emission of harmonic oscillations
according to IEC 61000-3-2
11301 DG
conforming to Class A
094129:
not applicable
Emission of voltage fluctuations /
flickers according to IEC 61000-3-3
11301 DG
conforming to Class A
094129:
not applicable
Anhang
Hinweise zur elektromagnetischen
Verträglichkeit (EMV) für
Ladegerät Modell 094129/11301 DG
Appendix
Information on the electromagnetic
compatibility (EMC) of
charger model 094129/11301 DG
Anexo
Indicaciones relativas a la compatibilidad
electromagnética (CEM) del cargador,
modelo 094129/11301 DG