Application
Instructions for use • C-MAC Monitor • NDQ566_EN_V1.0_02-2021_IFU_CE-MDR
31
Level 10
Toggle Freeze/Still image capture
Toggles between
Freeze with still image capture
and
Still image capture
.
Freeze with still image capture
freezes the live image on the screen and saves
the image to the SD card.
Still image capture
saves the live image on the screen to the SD card.
The setting is saved for this endoscope/camera.
This function is only activated if the endoscope/camera connected supports this
function.
Menu wheel
Activates the functions of the next level. You can move through several levels by
repeatedly pressing the menu wheel.
Wrench
Back to level
1
.
6.4 Transferring data from an SD card to a USB stick
The USB port serves exclusively to transfer data from the SD card to the USB stick. Data
cannot be saved directly to a connected USB stick.
ü
Before use with the C-MAC monitor, the USB stick must be formatted on the PC with
FAT (standard) and has only one partition.
Connect the USB stick to the C-MAC monitor.
ð
The video data are being transferred from the SD card to the USB stick.
When data are transferred from an SD card to a USB stick, the data are copied from the SD
card and not deleted. The data on the SD card can only be deleted manually on the C-MAC
monitor.
In the case of large video files or multiple recordings, the data transfer can take a long time.
A USB stick with a large storage capacity can delay the storage of images.
6.5 Messages
Messages indicating a critical status are shown on the monitor in two different ways.
Status message
–
The corresponding symbol in the top status bar has a white background.
Warning message