Safety
Instructions for use • C-MAC Monitor • NDQ566_EN_V1.0_02-2021_IFU_CE-MDR
9
or RFID) emit electromagnetic radiation, the function of the product may be impaired. High-
frequency communication equipment can affect medical electrical products and impair their
performance.
Do not use the product in the vicinity of a magnetic resonance tomograph (MRT).
Do not use the product next to or together with other devices. If such use is required,
monitor the product and the other devices, and follow the relevant instructions for use in
the event of malfunctions.
Portable RF communications equipment including peripheral devices (e.g., antenna cables
and external antennas) should be used no closer than 30 cm from the product, including
cables specified by the manufacturer.
Observe the information on electromagnetic compatibility; see chapter Electromagnetic
compatibility [p. 39].
In case of uncertainties, seek expert advice from KARL STORZ.
The use of accessories and cables other than those specified in the instructions for use may
result in increased emissions or decreased immunity of the product. When using other
accessories and cables, the operator is responsible for checking compliance with IEC
60601-1-2 for this particular product.
To prevent increased electromagnetic emissions or reduced electromagnetic immunity of
the product, only use accessories, transducers, and cables recommended or supplied by
the manufacturer.