ESR-Auto Plus

®

 RAM Reset Procedure 

For Model 504-V9,V10 and Model 505-V3 
Sep 2014, 880004-4

 

 

Page 2 of 4 

 
14. Instrument reads:  “

1.2ml (PAED)?”

   

 

The available tube settings are “1.2ml” only, “2.0ml” only or “Both 1.2ml & 2.0ml”. 

 

Press the red 

X

 button to scroll through the available options. 

 

Press the green 

 button to accept the desired tube type. 

 
15. Instrument reads:  “

Set Calibr.?

”   

 

Press the green 

 button.   

 

If the instrument reads “+0%”, press the green 

 button.   

 

If not, press the red 

X

 button, enter “0” and press the green 

 button. 

 
16. Instrument reads:  “

Set Meas. Time?

”   

 

Press the red 

X

 button.  

 
17. Instrument reads:  “

Audible Alarm”

 (Model 505 only)  

 

Press the green 

 button to access this feature. 

 

Press the green 

 button to enable the audible alarm. 

   
18. Instrument reads:  “

Set Printouts?”

   

 

Press the green 

 button.   

 
19. Instrument reads:  “

Auto Print?

”   

 

Press the green 

 button. 

 
20. Instrument reads:   “

Ticket Style?

”   

 

Patient results can print in either ticket style or log style.   

 

For ticket style, press the green 

 button.   

 

For log style, press the red 

X

 button. 

 

NOTE: If the instrument is interfaced with an LIS, the print style selected 

 

should match the data output format written in the LIS interface.   

 

 

21. Instrument reads:   “

Config Ports

?”   

 

Press the green 

 button. 

 
22. Instrument reads:   “

Int Prn On

?”   

 

Press the green 

 button. 

 
23. Instrument reads:   “

Com 1 On

?”   

 

Press the green 

 button. 

 
24. Instrument reads:   “

Com 2 On

?”   

 

Press the green 

 button. 

 
25. Instrument reads:   “

Main Printer, Int. Prn

?”   

 

Press the green 

 button. 

 
26. Instrument reads:   “

Print Setup

?”   

 

Press the red 

X

 button. 

 
27. Instrument reads:   “

Set Access Code

?”   

 

Press the red 

X

 button. 

Summary of Contents for ESR-Auto Plus 504-V9

Page 1: ...d X button 5 Instrument reads The software version will display followed by New Software OK for Setup Press the green button 6 Instrument reads Motor Setup Please Wait The instrument will cycle and print out the system settings This step may take a couple of minutes 7 Instrument reads IR Auto Set Remove Tubes OK Remove all tubes test rack and press the green button 8 Instrument reads Calibrating T...

Page 2: ... green button to enable the audible alarm 18 Instrument reads Set Printouts Press the green button 19 Instrument reads Auto Print Press the green button 20 Instrument reads Ticket Style Patient results can print in either ticket style or log style For ticket style press the green button For log style press the red X button NOTE If the instrument is interfaced with an LIS the print style selected s...

Page 3: ... Set IR Power Press the red X button 35 Instrument reads LCD Contrast Press the red X button 36 Instrument reads Set Alarms Press the green button 37 Instrument reads Over Fill If the instrument reads 690 press the green button If not press the red X button enter 690 and press the green button 38 Instrument reads 1 2 ml Low Fill If the instrument reads 510 press the green button If not press the r...

Page 4: ... Insert the test rack and press the green button The instrument will scan the rack twice and print the results 47 Instrument reads Set Thermometer Press the green button 48 Instrument reads Compensation On Press the green button 49 Instrument reads Celsius ___ Press the green button NOTE This setting is the internal working temperature of the instrument The temperature sensor reading will flash on...

Reviews: