5. PATIENT SAMPLE TRANSMISSION RESULTS
5.1 TICKET STYLE SAMPLE TRANSMISSION RESULTS
Note: An accurate date and time must be in ESR-Auto Plus in order for transmission of data to take
place. The ‘Auto Print’ selection in the User configuration must also be turned on for transmission of
data to take place.
Note: Results will vary for each ESR result (i.e. “ESR: 1 MM WESTERGREN(QM PAED TUBE”). The
sample transmission below has actual ESR results from 1 to 4, however, the reportable range of the
instrument is only 0-120. If the reading is above the reportable range a result of “>120” will be reported.
Note: The patient ID section of the ticket style transmission can have a maximum of 13 characters both
numerical and alphabetical. (i.e. “PATIENT ID:1234567890123”)
EXAMPLE OF ESR SAMPLE RESULT IN TICKET STYLE:
*********
*** ESR RESULT ****************
2009-02-18 12:04 POS:A E0000 S0353
PATIENT NAME:_______________________
PATIENT ID :1
ESR : 0 MM WESTERGREN(QM PAED TUBE)
*
TICKET STYLE TRANSMISSION FORMAT:
************ ESR RESULT ****************
2009-02-18 12:04 POS:A E0000 S0353
PATIENT NAME:_______________________
PATIENT ID :1
ESR : 1 MM WESTERGREN(QM PAED TUBE)
*
TICKET STYLE HEX RESULTS
0A 2A 0A
0A
44 20 54 55 42 45 29 0A
4E 28 51 4D 20 50 41 45
45 53 54 45 52 47 52 45
DATE
TIME
SAMPLE POSITION
ERROR
CODE
STATUS CODE
PATIENT ID NUMBER
ESR RESULT