ESR-Auto Plus Operator’s Manual 7
Chapter 2
|
Sample Preparation & Analysis
2.1 ESR-Vacuum Tubes
The ESR-Auto Plus is intended for use with Streck 1.2ml
ESR-Vacuum Tubes. ESR-Vacuum Tubes are also available
in High Altitude and Safety Coated 1.2ml configurations.
2.0ml ESR-Vacuum Tubes may be used by selecting the
alternate tube size in the User Configuration Menu.
All Streck ESR-Vacuum Tubes preserve the integrity of the
patient sample for up to 72 hours from the time of blood
collection when transported or stored at 2-10° C, or up to 4
hours at 18-30° C.
Follow the Clinical Laboratory Standards Institute (CLSI)
Procedures for the Erythrocyte Sedimentation Rate Test;
Approved Standard
, CLSI Document H02, for stability of
specimens collected in EDTA tubes.
2.2 Sample Collection
Patient samples may be drawn directly into the Streck
ESR-Vacuum Tube or transferred from an EDTA vacuum
tube. When using an EDTA tube, transfer the well-mixed
patient sample into the Streck ESR-Vacuum Tube. Do not
remove the sodium citrate in the ESR-Vacuum Tube. Fill the
ESR-Vacuum Tube to the fill line.
ESR-Vacuum Tubes contain liquid anticoagulant and must
be thoroughly mixed with the blood sample immediately after
the tube has filled. Failure to mix the sample completely and
immediately may result in the formation of microscopic blood
clots/aggregates that could alter results. Proper collection
and preparation of the ESR sample is critical to obtain
accurate results. Inaccurate results are most often a result
of improper sample handling.
CAUTION! Handle all biological samples and
blood collection sharps according to your
standard laboratory procedures. Seek medical
attention in the case of exposure to biological
specimens.
Follow CLSI
Procedures for the Collection of Diagnostic
Blood Specimens by Venipuncture
;
Approved Standard
,
CLSI document GP41, for guidance with sample handling.
Summary of Contents for ESR-Auto Plus 506
Page 1: ...ESR Auto Plus Operator s Manual Model 506 v 1 0 2 0 ...
Page 2: ......
Page 6: ......
Page 7: ...1 Introduction ...
Page 8: ......
Page 12: ......
Page 13: ...2 Sample Preparation Analysis ...
Page 14: ......
Page 20: ......
Page 21: ...3 Keypad Menu Overview ...
Page 22: ......
Page 24: ......
Page 26: ......
Page 36: ......
Page 38: ......
Page 44: ......
Page 51: ...7 Background of the ESR Test ...
Page 52: ......
Page 56: ......
Page 57: ...8 Performance Technical Specifications ...
Page 58: ......
Page 65: ...9 Safety Precautions ...
Page 66: ......
Page 69: ...10 Troubleshooting ...
Page 70: ......
Page 75: ...11 Quick Reference Guide ...
Page 76: ......
Page 80: ......
Page 82: ......
Page 84: ......
Page 85: ......
Page 86: ......