The Secotom-10 has a built-in cooling fluid system. The fluid coming
from the nozzles passes over the cut-off wheel and collects in the
bottom of the cutting chamber; where it then returns to the bath,
which is located under the cutting chamber. Access to this bath, for
cleaning and maintenance purposes, is possible by opening the front
cover of the Secotom-10.
Open and remove the front cover from the Secotom-10.
Lift the inlet tube (as illustrated).
Slide the fluid bath out gently.
Empty any old/used fluid from the bath and ensure that the bath
is clean (free from cutting debris).
Fill the bath with 3.69 litres of water and 110 ml Struers additive
for cooling fluid (Corrozip).
Slide the fluid bath back into position.
Check that the inlet tube is placed correctly so that the pump is
supplied with water.
Filling the Cut-off Machine
with Cooling Fluid
IMPORTANT
Always ensure that there is sufficient water in the tank as the recirculation
pump will be damaged if it is run dry.
Summary of Contents for Secotom-10
Page 1: ...Manual No 15037001 Date of Release 0 2010 Secotom 10 Instruction Manual...
Page 18: ...Secotom 10 Instruction Manual 14 2 Basic Operation Power Spindle speed 300 500 rpm Front Panel...
Page 49: ...Secotom 10 Gebrauchsanweisung Handbuch Nr 15037001 Auslieferungsdatum 1 0 2010...
Page 54: ......
Page 101: ...Secotom 10 Mode d emploi Mode d emploi no 15037001 Date de parution 1 0 2010...
Page 106: ......
Page 120: ...Secotom 10 Mode d emploi 14 2 Fonction de base Power Spindle speed 300 500 rpm Panneau frontal...
Page 151: ...Manual No 15037001 Date of Release 10 05 2010 Secotom 10 Spare Parts and Diagrams...
Page 158: ......
Page 159: ......
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 168: ......
Page 169: ......
Page 171: ......
Page 172: ......
Page 173: ......