Instruction Manual
19
Remove the cap covering the support hole placed on the top of
the TegraPol cabinet.
Remove the two caps covering the fastening screws on the left-
hand side of the TegraPol cabinet by lifting them off with a knife.
Lead the supporting column of TegraForce-1 down into the hole
at the back of the TegraPol.
Then both the distance between disc and specimen mover plate
and the position of the specimen mover plate over the disc have
to be adjusted.
To adjust the correct distance mount a preparation disc on
TegraPol and a specimen mover disc on the TegraForce-1.
Loosen the 2 fastening screws and adjust the space between the
disc and specimen mover plate to about 1 -1.5 mm.
Re-tighten the 2 fastening screws again.
To set the correct position, see Disc Margin in the following
section.
Setting up TegraForce-1 on
TegraPol
Support Hole
Fastening
Screws