To ensure optimal cooling, Unitom can be fitted with a Cooli unit. The
Cooling System 5 is a Cooli configuration designed for use with
Unitom.
To connect the Unitom to a cooling unit:
Plug the Cooli control unit’s communication cable into the
Unitom’s control socket.
Connect the quick coupling on the Unitom’s cooling fluid inlet
tube to the cooling unit’s pump outlet (
A
).
Insert the drain pipe
in the water outlet on the back of
Unitom-2 and mount the 90° elbow pipe
. Lubricate the sealing
ring with grease or soap to facilitate insertion. (Use the other
elbow pipe if more suitable).
Strip the steel spring from approx. 3 cm of the outlet hose and
cut. Bend the cut end towards the centre of the hose. Mount the
outlet hose
onto the elbow pipe and clamp the stripped
section using a hose clamp.
Check that the outlet hose slopes downwards when connected. If
necessary adjust the length of the hose.
Insert the open end of the hose into mounting hole in the bracket
on top of the Cooli filter unit (
B
).
Connect the cooling unit to the mains power supply.
Connecting a Cooli
Cooling Unit
Note
:
Cooling System 5 includes a static filter.
For intensive use, and for materials generating a lot of swarf,
a bandfilter is recommended.
Note
Before connecting the cooling unit to the Unitom,
follow the instructions in the Cooli Instruction Manual
to prepare it for use.
IMPORTANT
Before connecting, check that the mains voltage corresponds to the voltage
stated on the type plate on the side of the machine.
A
B
Summary of Contents for Unitom-2
Page 1: ...Manual No 15377001 Date of Release 20 Unitom 2 Instruction Manual...
Page 4: ......
Page 30: ...Unitom 2 Gebrauchsanweisung Handbuch Nr 15377001 Auslieferungsdatum 20...
Page 33: ......
Page 59: ...Unitom 2 Mode d emploi Mode d emploi No 15377001 Date de parution 20...
Page 62: ......
Page 89: ...Manual No 15377001 Date of Release 0 20 Unitom 2 Spare Parts and Diagrams...