Cleaning and Maintaining the 1088 Camera
29
Using Sterile Drapes
Using sterile drapes will ensure maximum longevity of your 1088 Camera
Head. For best results, follow the instructions provided by the drape
manufacturer. Sterile camera drapes for use with the 1088 Camera Head
can be ordered under Stryker product number 240-010-200.
Maintaining the 1088 Camera
Replacing the Fuses
1. Unplug the power cord from the wall outlet and remove the cord
from the camera console.
2. Unlatch the fuse holder above the AC inlet and remove it. (You
may need to press the tab on the fuse holder with a slender
screwdriver to release the latch.)
3. Replace the fuse with the same value and rating.
Warning To avoid the risk of fire, only use fuses of the value
specified on the fuse label located on the rear panel of
the camera control unit.
4. Reinstall the fuse holder until the tab snaps in place.
Periodic Maintenance Schedule
Warning To ensure safe operation of the Model 1088 Medical
Video Camera you should perform the following
procedure periodically:
Every 12 months, check the earth leakage current to <500µA (<300µA in
U.S.A.), ground impedance to <0.1 ohms, and power consumption less
than or equal to rated power. Use a true RMS digital multimeter and
safety analyzer to perform this test.
Note
Refer calibration and operating difficulties not detailed in
this manual to your Stryker Endoscopy sales
representative.
Disposing of the 1088
The 1088 must be disposed of according to local laws and hospital
practices. The device does not contain any hazardous materials.