16
Caution
Do not abruptly interrupt power to the unit. This will
turn off the fan and may cause severe damage to the
internal cooling system.
Checking the ESST Feature
The X7000 is equipped with Electronic Scope Sensing Technology
(ESST), a special safety feature that helps prevent accidental burns
caused by an unattended light cable. When operated with an ESST light
cable, the X7000 senses when the scope and light cable are separated
and places the light source in STANDBY mode. In STANDBY mode, the
X7000 will reduce light output to a minimum, preventing the light cable
from generating excessive heat.
To verify the ESST feature is active, perform the following test before
every surgical procedure:
1. Set up the X7000 system with an ESST light cable and scope,
and then power on the system.
2. Place the X7000 in RUN mode.
3. Remove the light cable from the ESST scope adapter.
The X7000 should return to STANDBY mode, indicating that the ESST
feature is functioning properly.
Warning If the unit fails to return to STANDBY mode, there may
be a fault with the ESST feature. In this case, do not
assume ESST safety protection.
Warning Even with ESST protection, never place the tip of the
light cable directly on drapes or on the patient as
burns or fire may result.
Using the X7000 with a Voice-Controlled System
Interface
The X7000 can be used in conjunction with Stryker voice-control systems
(Sidne™). For more information about using the X7000 with Stryker
voice-control systems, refer to the
Sidne™ Operating and Maintenance
Manual
(p/n 1000-400-653).