Installing the wall plate (Continued)
2.
Using the dimensions shown in (
), locate and drill the points where the floor plates will
attach.
Note:
When installing the wall plate for the Stryker Model 6371 cot fastener with the long rail assembly option
(6362-021-000), measure 33 in. for the horizontal dimension and 4 1/16 in. for the vertical dimension. The Stryker
Model 6371 cot fastener with the long rail assembly option is only compatible with the extended head section
assembly (9996-390-200).
24.000
(Ref.)
(61 cm)
6-1/8”
(15,6
cm)
14-15/16”
(37,9 cm)
80.000 (Ref.)
(203 cm)
2-5/8”
(6,6 cm)
14-5/8”
(37,1 cm)
1-1/8”
(2,9 cm)
3-5/8””
(9,2 cm)
37-3/8”
(94,9 cm)
6-1/2” (16,5 cm)
4-5/16” (10,9 cm)
A
B
C
D
E
F
G
H
Figure 1-15: Wall plate installation locations
A
Wall of vehicle
E
Cot perimeter
B
Antler floor plates
F
Cot head end
C
Rear of ambulance
G
Wall mounting bracket
D
Cot center line
H
Floor of vehicle
3.
Cut two 1 in. x 2-1/2 in. x 9/16 in. deep recessed openings (A) into the floor centered over the locating holes
).