Troubleshooting
WARNING:
DO NOT disassemble or service this equipment.
NOTE:
For service, contact your Stryker sales representative or call Stryker customer service. Outside the US,
contact your nearest Stryker subsidiary.
PROBLEM
CAUSE
ACTION
The handpiece
does not operate or
operates at a reduced
speed.
The battery pack is discharged.
Use a Stryker battery charger to recharge
the battery pack.
The battery pack is expended.
Replace the battery pack.
The function switch is in the safe mode
position.
Slide the function switch to the forward or
forward/reverse/oscillate position.
The drivetrain is malfunctioning.
Return the handpiece to Stryker for repair.
The handpiece
operates but the
accessory does not
move.
The attachment is not fully installed in the
handpiece.
Remove and install the attachment. Make
sure the attachment is securely locked in the
handpiece.
The shift collar is positioned between the
drill and ream mode positions.
Fully rotate the shift collar to either the drill
mode or ream mode position.
Excessive pressure is being applied to the
cutting accessory.
Release the pressure and allow the cutting
accessory to do the cutting.
The drivetrain is malfunctioning.
Return the handpiece to Stryker for repair.
The handpiece
continues to operate
after the trigger is
released.
The trigger is malfunctioning.
Depress the battery latch and slide the
battery pack out of the handpiece. Return
the handpiece to Stryker for repair.
The battery pack
becomes unusually
hot during use.
The circuitry is malfunctioning.
Use a Stryker battery charger to check
the integrity of the battery pack. See the
instructions for use supplied with the battery
charger for more information. Replace the
battery pack if required.