7205-001-700 Rev-AA
EN
www.stryker.com
19
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic environment - guidance
Radiated RF
IEC 61000-4-3
9 V/m
710, 745, 780,
5240, 5500,
5785 MHz, pulse
modulation 217
Hz, Maximum
power = 0.2 W
28 V/m
810, 870, 930
MHz, pulse
modulation 18 Hz,
Maximum power
= 2 W
28 V/m
1720, 1845, 1970,
2450 MHz, pulse
modulation 217
Hz, Maximum
power = 2 W
9 V/m
710, 745, 780,
5240, 5500,
5785 MHz, pulse
modulation 217
Hz, Maximum
power = 0.2 W
28 V/m
810, 870, 930
MHz, pulse
modulation 18 Hz,
Maximum power
= 2 W
28 V/m
1720, 1845, 1970,
2450 MHz, pulse
modulation 217
Hz, Maximum
power = 2 W
IEC 60601-1-2 4th Edition:
WARNING:
Portable RF equipment
(including peripherals such as antenna
cables and external antennas) should be
used no closer than 30 cm (12 inches)
to any part of the System 7 Dual Trigger
Rotary Handpiece (REF 7205-000-
000), including cables specified by the
manufacturer. Otherwise, degradation of
the performance of this equipment could
result.
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in each
frequency range.
b
Interference may occur in the
vicinity of equipment marked with the following
symbol:
(Non-ionizing electromagnetic radiation)
NOTE 1: At 80 MHz and 800MHz the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and
people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters,
an electromagnetic site survey should be considered. If the measured field strength in the location in which the System 7 Dual Trigger Rotary Handpiece
(REF 7205-000-000) is used exceeds the applicable RF compliance level above, the System 7 Dual Trigger Rotary Handpiece (REF 7205-000-000) should
be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating
System 7 Dual Trigger Rotary Handpiece (REF 7205-000-000).
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.