Guidance and manufacturer's declaration - electromagnetic immunity
The SwitchPoint Infinity
®
2 Control System is intended for use in the electromagnetic environment
specified below. The customer or the user of the SwitchPoint Infinity
®
2 Control System should assure
that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+/- 6 kV contact
+/- 8 kV air
+/- 6 kV contact
+/- 8 kV air
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Electrical fast
transient/burst
IEC61000-4-5
+/- 2 kV for power
supply lines
+/- 1 kV for input/output
lines
+/- 2 kV for power
supply lines
+/- 1 kV for input/output
lines
Mains power quality should
be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
+/- 1 kV differential
mode
+/- 2 kV common mode
+/- 1 kV differential
mode
+/- 2 kV common mode
Mains power quality should
be that of a typical
commercial or hospital
environment.
Voltage dips,
short
interruptions and
voltage variations
on power supply
input lines.
IEC 61000-4-11
<40% UT
(>60% dip in UT)
for 5 cycles
<70% UT
(>30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in UT)
for 5 sec
<40% UT
(>60% dip in UT)
for 5 cycles
<70% UT
(>30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in UT)
for 5 sec
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of the
SwitchPoint Infinity
®
2 Control
System requires continued
operation during power mains
interruptions, it is
recommended that the
SwitchPoint Infinity
®
2 Control
System be powered from an
uninterruptible power supply
or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Note UT is the a.c. mains voltage prior to application of the test level.