EMC Precautions
This device is considered medical electrical equipment and requires special precautions regarding EMC and
needs to be installed and put into service according to the information provided.
Portable and mobile RF communications equipment can affect this device’s performance and must be used in
accordance with the following information.
Required Equipment
This device must be used with the provided medical grade isolation transformer and power cord. Additionally
shielded video cables are required in order to assure proper function.
Using accessories or cables other than those approved by Stryker may result in increased emissions or decreased
immunity of the equipment or system.
Guidance and manufacturer's declaration - electromagnetic emissions
The SwitchPoint Infinity
®
2 Control System is intended for use in the electromagnetic environment
specified below. The customer or the user of the SwitchPoint Infinity
®
2 Control System should assure
that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The SwitchPoint Infinity
®
2 Control System uses RF energy
only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class A
The SwitchPoint Infinity
®
2 Control System is suitable for use
in all establishments other than domestic and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies