3. Set the top strap to the outer side of the respi-
rator and while holding the both ends of the
respirator with both hands, put the respirator
over the nose and mouth.
4. Hold the respirator with one hand, then put
the top strap over and settle it at top back of
your head with the other hand.
5. Pull the adjuster for the top strap, if neces-
sary, to adjust the length of the strap so that
the respirator can be attached firmly to your
face.
6. Adjust the tensions of the top and bottom
straps so that the respirator fits securely to
your face.
7. Press the nose clip with fingers of both
hands so that the respirator can be attached
firmly to your nose.
CAUTION
Do not hold both the top and bottom straps at the
same time when donning; otherwise, your face may
get injured by a corner of the respirator.
<Removing the respirator>
After removing the top strap over the back of your head, remove the bottom
strap hanging around your neck.
CAUTION
In order to prevent the secondary infection, special
care must be taken not to touch the surface of the res
-
pirator when taking off the respirator.
■
Proper fitting
In order to achieve the original performance of the respirator, the respirator must
have best fit. The user is required to be fit tested according to OSHA 29 CFR
1910.134 either qualitatively or quantitatively prior to use. Conduct a fit check
following this procedure every time you put on the respirator.
1. Wear the respirator in the same manner as
you actually conduct during work.
2. Cover the respirator with two hands and exhale.
If it is not possible to cover the respirator
with both hands, use a polyethylene sheet
to cover the surface of the respirator, hold-
ing it down with both hands and exhale.
3. Check that no air leaks through respirator-to-
face seal when you exhale. If you feel air leak-
age, adjust the position of respirator and/or tention of straps until the air leakage stops.
WARNING
1.
Before using the respirator in an actual operation, be sure to con
-
duct the inspection whether the respirator is properly fitted or not.
2.
If it is not possible to ascertain that the respirator is properly fitted,
do not use the respirator in an actual operation. Make the neces-
sary adjustments so that the respirator firmly fits your face.
If it is still not possible to ascertain that the respirator
is attached properly after those adjustments, inspect
each and every portion of the respirator at a safe
place.
(Please refer to “Check before donning”.)
<For your reference>
If your face is too small to fit properly to your chin, it is recommended to put the
bottom strap over the back of your head as shown in the figure below so that the
respirator is tightly fitted around the chin.
■
Storage Method
Store the respirator in a clean, cool, dark and dry place.
CAUTION
Do not pile up or bend the respirators to store ; it may
cause problems such as fissures, deformation, etc.
■
Disposal Method
Put the used respirator in an airtight bag to prevent the particulate matters from
scattering before disposal in accordance with the applicable laws and regula-
tions.
■
NIOSH Approval Label
1-26-1, Nishigahara, Kita-ku, Tokyo
114-0024, JAPAN
+81-3-6903-7525
THIS RESPIRATOR IS APPROVED ONLY IN THE FOLLOWING CONFIGURATION:
TC-
Protection
1
Respirator
Cautions and Limitations
2
DD02-N95+S
84A-
9373
Surgical/N95
X
ABCJMNOS
1. Protection
N95 – Particulate Filter (95% filter efficiency level) effective against
particulate aerosols free of oil; time use restrictions may apply.
2.
Cautions and Limitations
A-
Not for use in atmospheres containing less than 19.5 % oxygen.
B-
Not for use in atmospheres immediately dangerous to life or health.
C-
Do not exceed maximum use concentrations established by regulatory
standards.
J-
Failure to properly use and maintain this product could result in injury or
death.
M-
All approved respirators shall be selected, fitted, used, and maintained in
accordance with MSHA, OSHA and other applicable regulations.
N-
Never substitute, modify, add, or omit parts. Use only exact replacement
parts in the configuration as specified by the manufacturer.
O-
Refer to User’s Instructions, and/or maintenance manuals for information on
use and maintenance of these respirators.
S-
Special or critical User Instructions and/or specific use limitations apply.
Refer to User Instructions before donning.
Top
strap
Adjuster
Nose
clip
For more information regarding the handling method of the
product, please contact:
Head O
ffi
ce
1-26-1, Nishigahara, Kita-ku, Tokyo 114-0024, Japan
Tel : +81-3-6903-7525