18
6.3 Standard Sterilization Program
A. Before start Sterilization program please refer to “6.2 Prepare Sterilization” section.
B. How to set the Standard Sterilization program
Figure 11
C. Press “STERI TEMP”
to select 121°C or 134°C
D. Press “STERI ITEM”
to select unwrapped instruments or wrapped items. The
corresponding parameters will be:
Sterilization Temperature
121°C
Sterilization Temperature
134°C
Unwrapped
22 min.
4 min.
Wrapped
30 min.
15 min.
E. Press Dry Timer Button
to select a maximum 40 min. drying time. (*If drying is
required.)
F. Press the START Button
, the setting temperature, duration, and dry time displayed two
times on the three-digit display, and then the sterilizer will automatically run through the
selected program. The current progress of the sterilization cycle is indicated by the
illuminated LED on the Process Status Indicator. (Figure 12).
Figure 12
121°C or 134°C
120°C
103°C
(Room Temp.)°C
0 bar
(kgf/cm
2
)
Vacuum