6.5 Liquid Program (optional, exclusive either to Re-Dry function)
A. Before start Sterilization program please refer to “6.2 Prepare Sterilization” section.
B. How to set the Liquid program
Figure 14
Caution:
After the sterilization step finished, the unit will take to exhaust
spontaneously. Then the COMPLETE indicator light is illuminated on the
Process Status Indicator.
C. Press
to start “Liquid Program”. The corresponding sterilization time will be:
Sterilization Time
40 min.
Sterilization Temperature
121°C
Dry Time
N/A
D. Press the START button
, the setting temperature, duration, and dry time displayed two
times on the three-digit display, and then the sterilizer will automatically run through the
selected program. The current progress of the sterilization cycle is indicated by the
illuminated LED on the Process Status Indicator.
Note:
The DRY step won’t be performed.
Figure 15
121°C
(Room Temp.)°C
0 bar
(kgf/cm
2
)