23
E. Open the door and take out the sterilized items.
Warning:
Check the Pressure Gauge is reading ZERO before opening the door.
Warning:
Beware of steam when opening door after a sterilization cycle.
Warning:
Be careful when removing the sterilized items as the metal surfaces might
still be hot. Always wear suitable hand protection to remove the box or use
the appropriate aids (tray holder) to lift the trays.
Warning:
If using the sterilizer continuously, it’s required to have a 20 min. interval
between each sterilization cycle to allow the unit to cool.
6.7 B.D. TEST Program
A. Before start Sterilization program please refer to “6.2 Prepare Sterilization” section.
B. How to set the B.D. TEST program
Figure 18
C. Press
to start “B.D. TEST Program”. The corresponding sterilization time will be:
Sterilization Time
3 min.
Sterilization Temperature
134°C
Dry Time
NA
D. Press the START Button
, the setting temperature, duration, and dry time displayed two
times on the three-digit display, and then the sterilizer will automatically run through the
selected program. The current progress of the sterilization cycle is indicated by the
134°C
120°C
103 °C
(Room Temp.)°C
0 bar
(kgf/cm
2
)
Vacuum