26
Warning:
Check the Pressure Gauge is reading ZERO before opening the door.
6.9 Vacuum Release
Press the Vacuum/Pressure Release Button
to release the chamber pressure. The
message “OP” will be displayed to notify operation. Please check the Pressure Gauge is reading
ZERO before opening the door
6.10 Reset / Stop
Press the RESET/STOP Button
3 seconds to suspend the current step of a sterilization
program.
Note:
You may press the Emergency Button
to release the chamber pressure.
Warning:
Check the Pressure Gauge is reading ZERO before opening the door.
6.11 Emergency Stop
Press the Emergency Button
3 seconds to stop the program and release the
pressure/vacuum. The sterilizer will sound to alert, and the Error message E07 will be displayed
to notify an emergency operation. Please wait till the pressure gauge is reading ZERO, or press
the Reset/Stop Button
to stop the error message and also to reset the program.
Warning:
The Emergency Button can only been pressed when there’s an unusual event or
emergency. The sterility of the sterilized items should be verified again.
Warning:
Disposal of the items which is sterilized by unfinished cycle should be in
accordance with the local laws. Do not handle them as general waste.
Note:
If the Emergency Button had been pressed without opening the door, you may require
repeating this emergency to release the pressure.
6.12 Placement for items to be sterilized
Please place items to be sterilized on the tray properly in order to have the best drying result.