47
9. Maintenance Instructions
Warning:
Failure to follow the Maintenance Instructions will adversely affect performance
and lifespan of the sterilizer, and may invalidate the warranty.
Warning:
Before conducting maintenance, please turn off the sterilizer and disconnect from
the power supply. Check the sterilizer has cooled down to room temperature.
Warning:
Make sure that pressure gauge is reading ZERO before opening the door.
Caution:
Before conducting maintenance, confirm that the chamber is empty without loads.
Correct and regular maintenance is required to optimize the performance of the sterilizer.
Failure to follow the Maintenance Instructions will adversely affect performance and lifespan of
the sterilizer.
9.1 Daily
-
Clean the external surfaces with soft cloth.
Note: Use only quaternary disinfectants to clean the units. Use of alcohol cleaner containing
substantial of alcohol in the formula may damage the faceplate.
-
Wipe the inside of the chamber, door and the gasket with a damp, lint-free cloth.
-
Check the water level. Top up with distilled water only.
-
Ensure the vent holes in the water reservoir cap (
Figure 6) are not blocked.
-
Check the status of the power cord. Call for service if breakage comes up.
9.2 Weekly
-
Clean the box, tray frame and trays with detergent, or a non-corrosive stainless steel
cleaner and water, using cloth or sponge.
-
Clean the filter use a wrench (or Knife Screwdriver) to unscrew the filter nut
counterclockwise as shown in Figure 51 to Figure 53.